Cr 2021-02-01 vs. 2.3.
User Manual for VCS-20Cr
25
Cleaning the device
.
The device can be cleaned using damp, soft wipes.
Cleaning the hoses:
Fill an empty, rinsed hydrogen peroxide bottle with distilled water. Place it to the device.
Press
Circulate
to clean the hoses. This should be done after treatment when it’s known
that the device is not being used for a while.
Transportation and storage
.
Store the device case closed at room temperature, protected
from excess moisture. Note that some of the hydrogen peroxide solution will remain
inside the feeding system.
Troubleshooting
The main switch doesn’t lighten up?
Make sure the device receives power.
Check fuse F1.
The process does not start?
-
Check the connection of the hydrogen peroxide sensor
-
Check the volume of the space, it can’t be 0
-
Hydrogen peroxide bottle is getting empty
-
Check fuses F2-F4.
Note!
If the same fuse blows again, the device should be delivered to a country-specific
authorized distributor or the manufacturer for service.
Heat plate stops heating?
In case that the heat plate stops heating, the unit must be delivered to a country-specific
authorized distributor or to the manufacturer. The device contains an overheating
protection that will be discharged if the surroundings temperature is too hot or the fan
isn’t working and the cell is overheated.
Disposal?
Before disposal the device it has to be decontaminated. Send the device to the
manufacturer for decontamination. The device can also be scrapped following local
regulations. Note that the device may contain traces of hydrogen peroxide
.
Cleamix VCS-20Cr bio-decontamination solution
Cleamix VCS-20Cr unit is a very powerful hydrogen peroxide vapour generator. It is
designed for effective bio-decontamination.
Appendix I