Model G14SD Instruction Manual
M0021_E
8
G14SD
Chapter 3 Operation
The G14SD is designed to run continuously to prevent any fume build up inside the unit. The red
light in the on/off switch will be illuminated when system is operating.
Chapter 4 System Maintenance
4.1 Changing the Filters
The GUS
System has a patented carbon filter that should last six months, even during continuous
use. The pleated design of the filters eliminates the need for a pre-filter.
NOTE
Contact Medivators Reprocessing Systems to order replacement filters.
CAUTION
The system’s filter must be replaced every six (6) months.
Wear personal protective equipment – such as eyewear, facemask,
moisture-resistant clothing and chemical-resistant gloves.
Disconnect the system’s power supply before removing the filter.
1. Turn the system’s power switch OFF, and disconnect the power cord from the wall outlet.
2. Unlatch the top lid.
3. Remove the old filter by pulling up on the strap located on top of the filter. Place in
plastic bag and discard with other non-hazardous waste.
4. Replace with a new filter. Verify that the direction of the arrows on the filter is pointing to
the back of the GUS
System.
5. Close and latch lid.
4.2 Cleaning the System
1. Turn the power switch OFF and disconnect the power cord from the wall outlet.
2. To clean metal surfaces: Use a clean, lint-free cloth moistened with a mild detergent
solution and water. A mild solvent may be used on these surfaces for difficult-to-clean
areas
3. Make sure that the unit is completely dry before use.