- 76 -
23. CONNECTING ADDITIONAL KEYS
It is possible to connect external tare and print buttons by special luster through port
RS232.
Printer or computer can be connected to the cluster.
Connected elements are not standard accessories of the balance.
1. balance
2. cable which connects balance with cluster
3. cluster
4. button
TARA
5. button
24. LIST OF COMMUNICATIONS COMPUTER-BALANCE
Function
INTERFACE RESET
Command
R CR LF (zero actual orders, restore factory setting)
Function
SEND ALL COMMANDS FROM THE BALANCE
Command
PC CR LF (all recorded information in commands in the balance programme
are sent from the balance)
Function
SEND THE RESULT IN BASIC UNIT
Command
S CR LF (result is sent from the balance in basic interval after stability)
Function
SEND RESULT IN BASIC UNIT IMMEDIATELY
Command
SI CR LF
Function
SEND THE RESULT IN ACTUAL INTERVAL
Command
SU CR LF (result in actual unit is sent from the balance after stability)
Function
SEND RESULT IN ACTUAL INTERVAL IMMEDIATELY
Command
SUI CR LF
Function
ZERO THE BALANCE
Command
Z CR LF (set the balance to zero after it reaches stability
Function
ZERO IMMEDIATELY
Command
ZI CR LF
Function
TARE WHEN STABLE
Command
T CR LF
Function
TARE THE BALANCE IMMEDIATELY
Command
TI CR LF