- 39 -
Draw. 30. Display with AKD function on
1- Single piece mass
2 - all elements mass
3 - AKD function(automatic correction of preciseness)
Display shows quantity of pieces which are on the pan (10 pieces). If less than counted
actually quantity is added mass of singular piece is corrected. In this case APW = 5.2282
to 5.1837. From this moment following pieces are counted in relation to singular mass.
This way mass of singular piece can be determined on base of batch standard.
The are four conditions of AKD (Automatic Correction of Preciseness) in the balance
software
1. quantity of pieces (after adding) must be higher than it was previously
2. quantity of pieces (after adding) must be less than twice quantity which was on
the display before adding
3. actual quantity must be in tolerance
±
0,3 of the total value,
4. the result must be stable.
If user decides that batch quantity is enough singular piece mass must be introduced into
the balance memory after pressing the key
RIGHT ARROW
.