Hardware installation
© 2021 Cisco and/or its affiliates. All rights reserved.
Page 31 of 177
Figure 7. Device connector ports
Connect a suitable power source to the unit by doing the following steps:
1. Only use a patch Ethernet cable to connect the power source and
the unit.
2. Insert the RJ45 connector leading from the power source into the
Ethernet port labelled
LAN1/POE
.
5.1.5. Rebooting the firmware and resetting the unit to factory
defaults
The Cisco FM3500 Endo hardware can be rebooted and reset to factory
default condition using the procedures in this section.
CAUTION
The unit is designed to take power through 48 Vdc
Power- over-
Ethernet
only, and does not have a dedicated power port. Do
not attempt to connect any other source of electrical power to
the unit.
The unit is designed to use an
IEEE 802.3at
PoE injector. Do
not connect any PoE injector conforming to
IEEE 802.3af
to the
unit. Such injectors have a higher voltage variance than the
design specification of the unit allows for, and may result in
unreliable performance.
IMPORTANT
The following procedure shows how to do a 'hard' (device
firmware) reboot. To do a 'soft' (device software) reboot, refer
to
“Resetting the unit to factory defaults” (page 138)