MAINS POWER
The generator is designed to operate off a single phase 220-240V 50 Hz
1
AC supply capable of
supplying a minimum of 300W. A suitable IEC mains lead is supplied with the unit.
WARNING
:
THIS APPLIANCE MUST BE EARTHED.
The wires in the mains lead are colour coded in accordance with the following code:
Green - and - yellow Earth
Blue : Neutral
Brown : Live
As the colours of the wires in the mains lead may not correspond with the coloured markings
identifying the terminals in your plug, proceed as follows:
The wire which is colour coded green-and-yellow must be connected to the terminals in the plug
which is marked with the letter 'E' or by the earth symbol which is ' ', or which is coloured green
or green-and-yellow.
The wire which is colour coded blue must be connected to the terminal which is marked with the
letter 'N' or coloured black.
The wire which is colour coded brown must be connected to the terminal which is marked with
the letter 'L', or coloured red.
MAINS FUSING
The mains power fuse is located underneath the IEC connector in a small carrier. Access is gained
by first removing the IEC power cord and then using a flat blade to unclip and withdraw the fuse
carrier. The fuse is rated as 20mm 250vAC T3.15A. and a suitable spare is also provided. If this
needs to be replaced it is essential for proper protection that it is replaced with one of the same
specification. Should the fuse continue to blow, then internal damage has occurred and the whole
unit must be returned to your dealer for service.
_____________________________________________________________________________
1
110v and 60Hz versions available to special order.