PROBLEMBEHANDLUNG

TROUBLE SHOOTING

EXTENDED PIVOT ARM LIGHT DUTY

DREIGELENKARM LIGHT DUTY

- 9 -

CIM med 

GmbH • Euro-Industriepark • Margot-Kalinke-Str. 9 • 80939 Munich • Germany         www.cim-med.com                                      

Problem

Problem caused by

Remedy

ME equipment does not appear level or parallel 
to floor

Channel not even/ straight

Adjust wall channel or reinstall channel.

Adapter and rear pivot part are not tightened 
completely.

Please check that there is no crack or movement 
between these parts. Tighten the screws and 
make sure that adapter and console are pulled 
tight with a torque moment of 25 Nm.

Adapter not parallel to wall channel: (4) cylinder 
bolts are pulled tight without having adjusted 
the adapter within the channel.

Remove cylinder bolts on both sides. Tighten 
first the cylinder pins on the left side. See 
assembly instruction adapter vertical/ horizontal 
wall channel.

Monitor adaption VESA / flat panel are not 
mounted parallel to floor.

Loosen the (4) M4 screws, adjust the monitor 
and pull the (4) screws tight again. Make sure 
that the monitor appears level or parallel to the 
floor.

Problem

Ursache des Problems

Problembehandlung

ME-Gerät nicht gerade bzw. parallel zum Boden

Wandkanal nicht gerade montiert

Wandkanal ausrichten oder neu montieren

Adapter nicht fest mit Konsole verbunden

Adapter aus der Wandschiene nehmen und 
die 2 x M8 Schrauben, an der Hinterseite des 
Adapters, mit einem Drehmoment von 25 Nm 
anziehen.

Adapter nicht parallel in dem Wandkanal 
montiert: (4) Zylinderstifte in der falschen 
Reihenfolge angezogen.

Lösen Sie Zylinderstifte auf beiden Seiten. 
Ziehen Sie als erstes die Zylinderstifte auf der 
linken Seite an. Siehe Montageanleitung Adapter 
vertikaler / horizontaler Wandkanal.

Monitorbefestigung VESA / Flachbildschirm 
schief auf der VESA-Montageplatte montiert.

Lösen Sie die 4 x M4 Schrauben, richten Sie den 
Monitor aus und ziehen Sie die (4) Schrauben 
wieder fest. Beachten Sie, dass der Monitor 
mit M4 Schrauben befestigt ist, die Aufnahme-
bohrung jedoch größer ist. Dadurch muss der 
Monitor exakt ausgerichtet werden.

Adapter mit 
Zylinderstiften

Konsole

TROUBLE SHOOTING

PROBLEMBEHANDLUNG

Adapter with 
cylinder bolts

Rear pivot part

Summary of Contents for 33030

Page 1: ...nhinweise Installation Bedienung und Wartung 2 Benötigtes Werkzeug 3 Montage von ME Geräten 3 Reinigung und Desinfektion 3 Wartung Art und Häufigkeit 3 Aufbewahrung 4 Bei Rückfragen 4 Name des Herstellers 4 CE Kennzeichen 4 Autorisierter Fachhändler 4 Lieferumfang 5 Monitoraufnahmen 6 Montageschritte 8 Problembehandlung 9 Wartungsplan 10 TABLE OF CONTENT Area of application Intended use 2 General ...

Page 2: ...allation und den bestimmungs gemäßen Gebrauch des Tragarmes sicher zu stellen Bitte überprüfen Sie das zugeladene Gewicht Dieses kann sich verändern wenn zusätzliches Zubehör zum eigentlichen Monitorgewicht hinzugefügt oder entfernt wird Stellen Sie sicher dass das maximal zulässige Gewicht nicht überschritten ist PLEASE READ THIS INSTRUCTION CAREFULLY BEFORE INSTALLING Clarification of non standa...

Page 3: ...r Tragarm ist ausgelegt für 26 000 Bewegungen oder 13 000 Zyklen ein Zyklus eine volle Bewegung bis zum Anschlag und zurück Bei 10 Bewe gungen pro Tag und 365 Tagen Bedienung an 7 Tagen in der Woche im Product modifications disassembling and reassembling and or mounting any of the components in any other way than described below are permitted only after written approval by the CIM med GmbH constru...

Page 4: ... kontaktieren Sie bitte unser Serviceteam Telefon 49 89 978 94 080 oder Ihren Händler vor Ort NAME DES HERSTELLERS CIM med GmbH Margot Kalinke Str 9 80939 München Deutschland CE KENNZEICHEN AUTORISIERTER FACHHÄNDLER week means 3 650 movements per year which correspond to a durability of 7 1 years In case of damage done to the load bearing parts of the system it has to be strictly refrained from fu...

Page 5: ...ANG SCOPE OF DELIVERY EXTENDED PIVOT ARM LIGHT DUTY DREIGELENKARM LIGHT DUTY 5 CIM med GmbH Euro Industriepark Margot Kalinke Str 9 80939 Munich Germany www cim med com Innensechskantschlüssel 4 0 Hex key 4 0 1x ...

Page 6: ...VICES For rear mounted devices see separate assembly instructions RÜCKSEITIGE GERÄTEANBINDUNGEN Für die Adaption rückseitig finden Sie separate Montageanleitungen VESA 75 100 Assembly Instruction Monitor Adaptation VESA VESA 75 100 Montageanleitung Monitoraufnahme VESA VESA 75 Assembly Instruction Monitor Adaptation VESA VESA 75 Montageanleitung Monitoraufnahme VESA VESA 75 QUICK RELEASE Assembly ...

Page 7: ...ntageplatte MINDRAY QUICK RELEASE Assembly Instruction Monitor Adaptation Mindray Quick Release MINDRAY QUICK RELEASE Montageanleitung Monitoraufnahme Mindray Quick Release PHILIPS TABLE TOP MOUNT Assembly Instruction Monitor Adaptation Philips Table Top Mount PHILIPS ZAPFENADAPTER Montageanleitung Monitoraufnahme Philips Zapfenadapter KEYBOARD TRAY Assembly Instruction Keyboard Tray for height ad...

Page 8: ... der Schraube 10 Nm kann das Gussteil bzw Gewinde brechen und zu Funktionsverlust führen ACHTUNG Beim Lösen beider Schrauben besteht die Gefahr des Herunterklappens des ME Gerätes GEFAHR DER FINGERQUETSCHUNG ASSEMBLY STEPS 1 Mount the support arm to its designated place wall channel tube DIN rail etc The illustration shows a wall channel adapter Please refer to the corresponding assembly instructi...

Page 9: ...pull the 4 screws tight again Make sure that the monitor appears level or parallel to the floor Problem Ursache des Problems Problembehandlung ME Gerät nicht gerade bzw parallel zum Boden Wandkanal nicht gerade montiert Wandkanal ausrichten oder neu montieren Adapter nicht fest mit Konsole verbunden Adapter aus der Wandschiene nehmen und die 2 x M8 Schrauben an der Hinterseite des Adapters mit ein...

Page 10: ...on der vorhandenen Rotationsanschläge Function of the existing rotational stop 2 6 Keine unzulässigen Veränderungen am Tragarm No unauthorised product modifications 2 7 Keine Verschmutzung No contamination 2 8 Keine Oberflächenbeschädigung oder Korrosion No damage or mark on the surface or corrosion 3 Prüfung mechanischer Teile Inspection of mechanical components 3 1 Spielfreier sicherer Sitz der ...

Page 11: ......

Page 12: ...CIM med GmbH Euro Industriepark Margot Kalinke Str 9 80939 Munich Germany www cim med com Assembly Instruction Extended Pivot Arm Light_2019 01_REV 30 09 2019_DE_EN ...

Reviews: