PROBLEM 5: AIR PRESSURE LEVEL IS SIGNIFICANTLY LOW NO MATTER HOW
HIGH THE AIRFLOW LEVEL IS CALIBRATED
Case 1:
Check the mains voltage (AC power source). If the voltage level falls
significantly low, about 15-20% lower than the standard, there will also be a
noticeable drop in the air pressure level.
SOLUTION:
Please refer to your local power service provider.
Case 2:
The microcontroller might have detected the operating frequency
incorrectly. The airflow level is noticeably weaker.
SOLUTION:
Try to press the "Reset" button on the panel and let the device re-detect the
proper operating frequency. Note that resetting the device will also reset ail previously
defined configurations.
PROBLEM 6: THE UNIT IS VIBRATING TOO MUCH
Check if the 4 screws that hold the pump in place are properly and tightly connected.
Unplug the system from the main power source before opening the case to check the
internal settings.
PROBLEM 9: AIR PRESSURE DROPS WHEN USING THE SUCTION PEN
DESCRIPTION:
The air level pressure drops when using the suction pen and does not
recover to its previous height in the air gauge.
SOLUTION:
Decrease the air pressure level, set air pressure level at about 80.
NOTE:
Air level may drop a little when the suction pen is picking up ICs. The air pressure
level gauge should revert back to its previous state when the IC is release.
Blockages in suction pen’s tube and nozzle may cause the air pressure to drop. Clean the air
passage for maximum air pressure capacity regularly.