8.20 TIME-Time Setting 

To set or change the system time. Place rotary switch at position “TIME” mode. The time and 

date on the third display panel should blink. Enter from left to right in the following format 
HH:MM:SS. Press On ( ), under ( ) button printed on the overlay in white word to set the 
system date. Press the “ENTER” button to confirm. Exit this function by pressing ESC button. 

Except these 2 positions mentioned above, all other position are for system setting, the indica-

tor “SET” shows at left side of LCD when place the rotary switch at any of these system setting 
positions. And 2nd functions of buttons are active. 
HH: Interval Hour (0~23) 
MM: Interval Minute (0~59) 
SS: Interval Second (0~59) 
MAX: 23:59:59 
MIN:00:00:01 

 
8.21 DATE-Date Setting 

To set or change the system data. Place rotary switch at position “Date” mode. The time and 

date on the third display panel should blink. Enter from left to right in the following format YY: 
MM: DD. Press On ( ), under ( ) button printed on the overlay in white word to set the sys-
tem date. Press the “ENTER” button to confirm. Exit this function by pressing ESC button. 

Except these 2 positions mentioned above, all other position are for system setting, the indica-

tor “SET” shows at left side of LCD when place the rotary switch at any of these system setting 
positions. And 2nd functions of buttons are active. 
Y: Interval year (20 "00" ~20 "99") 
M: Interval month (1~12) 
D: Interval date (1~31) 
MAX: 99:12:31 
MIN: 00:01:01

 

9. Maintenance 

Periodically wipe the case with a damp cloth, detergent and do not use abrasives or solvents.

 

 

 

WARNING: 

To avoid possible electrical shock, disconnect the 

thermocouple connectors from the thermometer 

before removing the cover.

 

 

18 

R/S-TYPE 

 

±(0.05% rdg + 2°C) on 0°C to 1767°C 
±(0.05% rdg + 4°F) on 32°F to 3212°F 

 

CHA (RTD) Measuring Range: 

Acceptable RTD Type: 

Pt385(100ohm)

Pt3916(100ohm)

Pt3926(100ohm). 

Temperature: 

-200°C to 800°C, -328°F to 1472°F 

Resolution: 

0.01, 0.1 for above  ±200 °C/°F. 

Accuracy: 

±0.05% rdg + 0.2°C or ±0.05% rdg + 0.4°F 

Resistance: 

0 ohm to 199.99 ohm, 200.0 ohm to 400.0 ohm. 

Resolution: 

0.01, 0.1 for above  200 ohm. 

Accuracy: 

±0.05% rdg + 0.1 ohm. 

 

CHB (ADAPTER) Measuring Range: 

 

Accuracy:

  ±(0.1% rdg + 10dgts) on 200mV 

 

±(0.5% rdg + 4°dgts) on 2000mV 

Stated accuracy at 23°C ± 5°C, <75% R.H.   

Temperature Coefficient: 

0.1 times the applicable accuracy specification per °C from 0°C to 

18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F). 

Input Protection: 

24V DC or 24V AC rms maximum input voltage on any combination of   

input pins. 
 

3. Names of Outer Parts: 

Front Side:   

1.

 

LCD. 

2.

 

Push buttons. 

3.

 

Rotary switch. 

4.

 

Socket of CHB (channel B) for voltage input. 

 

 

Summary of Contents for 506A

Page 1: ...OPERATOR S MANUAL MODEL 506A...

Page 2: ...r ID 001 Ex PC send 001L00006 r then meter response 001L0000111554170704111220420900115000001 r PC sends 001D00006 r then meter will dump all of data of group 6th 001D00006 rxxxxxx rxxxxxx rxxxxxx r M...

Page 3: ...of Buttons 9 8 1 D1 Setting the Main Display 9 8 2 D2 Setting Second Display 10 8 3 D3 Setting Third Display 10 8 4 T1 TYPE and T2 TYPE Setting the Types of Thermocouple for T1 or T2 10 8 5 T1 ADJ an...

Page 4: ...s 5 Alarm Counter AC 5 S Counts 5 Saved Data SD mode 1 mode 2 34 S GroupID SC c f mode T1_type T2_type Time Value r 5 5 1 1 1 1 12 6 2 Saved Data SD mode 3 34 S GroupID SC c f dp cha_type Time Value r...

Page 5: ...trument contains no operator serviceable part Service should be done by qualified person only Do not use the meter if the meter or test leads look damaged or if you suspect that the meter is not opera...

Page 6: ...S TYPE Resolution 1 0 C to 1767 C 32 F to 3212 F N TYPE Resolution 0 1 50 C to 1300 C 58 F to 2372 F Resolution 0 1 C 1 C 0 1 F 1 F Auto Ranging 2 2 Accuracy K J T E TYPE 0 05 rdg 0 3 C on 50 C to 13...

Page 7: ...the indica tor SET shows at left side of LCD when place the rotary switch at any of these system setting positions And 2nd functions of buttons are active Y Interval year 20 00 20 99 M Interval month...

Page 8: ...function APO Function will be Disabled also in Follow situation External power is used RS 232 function is used MAX MIN MAX MIN AVG function is used Data logger function is used 8 18 INVT Set Time inte...

Page 9: ...e blinking position with and key Press ESC key to end the set up mode indicator C or F ends blinking and indicator SET disappears but keep the previous setting unchanged Press CANCEL key to return to...

Page 10: ...ory Clear function or to leave this function by turning the Rotary Switch to an other position SET SET function Select the channels for data logging When the blinking indicator shows SET press key to...

Page 11: ...on marked with turning the rotary switch to any other position automatically turns on Model 506A5 Num bers and marks appear in the LCD display When APO Auto Power Off is enabled refer to 8 17 for enab...

Page 12: ...refer to section 8 18 To read the logged data please refer to section 8 12 8 8 11 CHB RANGE Select the Resolution for Cannel B CHB Channel B can accept the voltage signal up to 2 volts above 2 volts...

Page 13: ...And not function when LOG function is set as multi channel logging During measuring press SAVE button to record current measurement reading of the channel shown in Main Display and time To read the r...

Page 14: ...are known the offset adjustments can be set to each of T1 and T2 for compensating the off set er ror 10 RTD TYPE 8 6 CHA TYPE Selecting the Types of RTD for CHA Besides accepting the RTD of type Pt 38...

Reviews: