Digital Power Meter 66203/66204 User’s Manual
6-4
6.2.3 Questionable Status
■
The Questionable Status registers inform you one or more questionable status
conditions which indicate certain errors or faults have occurred to at least one channel.
Table 6-2 lists the questionable status conditions that are applied to the electronic load.
These conditions are same as the channel status conditions. Refer to Table 6-1 for a
complete description.
■
When a corresponding bit of Questionable Status Condition register is set, it indicates
the condition is true.
■
Program the PTR/NTR filter to select the way of condition transition in the Questionable
Status Condition register that will be set in the Event registers.
■
Reading the Questionable Status Event register will reset it to zero.
■
The Questionable status Enable register can be programmed to specify the
questionable status event bit that is logically ORed to become Bit 3 (QUES bit) in the
Status Byte register.
Table 6-2 Bit Description of Questionable Status
Mnemonic
Bit
Value
Meaning
OVR
0 1
Over
voltage
range.
OCR
1
2
Over current range.
OCP
2 4
Over
current.
RCE
3
8
Integration range changed error.
RCE
4
16
Inrush range changed error.
RCE
5
32
Energy range changed error.
6.2.4 Output Queue
■
The Output Queue stores output messages until they are read from the electronic load.
■
The Output Queue stores messages sequentially on a FIFO (First-In, First-Out) basis.
■
It sets to 4 (MAV bit) in the Status Byte register when there are data in the queue.
6.2.5 Standard Event Status
■
All programming errors that have occurred will set one or more error bits in the Standard
Event Status register. Table 6-3 describes the standard events that apply to the
electronic load.
■
Reading the Standard Event Status register will reset it to zero.
■
The Standard Event Enable register can be programmed to specify the standard event
bit that is logically ORed to become Bit 5 (ESB bit) in the Status Byte register.
Table 6-3 Bit Description of Standard Event Status
Mnemonic Bit Value
Meaning
OPC
0 1
Operation Complete.
This event bit generated is responding to
the *OPC command. It indicates that the device has completed
all of the selected pending operations.
QYE
2 4
Query Error
. The output queue was read when no data were
present or the data in the queue were lost.
DDE
3 8
Device Dependent Error
.
Memory was lost, or self-test failed.
Summary of Contents for 66203
Page 1: ...Digital Power Meter 66203 66204 User s Manual...
Page 2: ......
Page 3: ...Digital Power Meter 66203 66204 User s Manual Version 1 0 March 2014...
Page 8: ...vi...
Page 12: ......
Page 22: ......
Page 26: ......
Page 76: ......
Page 126: ......
Page 130: ......
Page 132: ......
Page 136: ......
Page 144: ......