Remote Operation
5-13
5.6.2.2
CONFIGURE
1.
CONFigure:BEEPer
Description:
Sets the beeper to ON or OFF.
Syntax:
CONFigure:BEEPer <Arg>
Parameter:
<Arg>: ON | OFF
Example:
CONF:BEEPer ON
Query Syntax:
CONFigure:BEEPer?
Return Parameter:
ON | OFF
Query Example:
CONFigure:BEEPer?
Return Example:
ON
2.
CONFigure:OUTPut
Description:
Sets the output voltage/current.
Syntax:
CONFigure:OUTPut <Arg>
Parameter:
<Arg>: ON | OFF
Example:
CONFigure:OUTPut ON
Query Syntax:
CONFigure:OUTPut?
Return Parameter:
ON | OFF
Query Example:
CONFigure:OUTPut?
Return Example:
ON
3.
CONFigure:FOLDback
Description:
Sets the type of FOLDBACK PROTECT.
Syntax:
CONFigure:FOLDback <Arg>
Parameter:
<Arg>: DISABLE | CVTOCC | CCTOCV
Example:
CONFigure:FOLD DISABLE
Query Syntax:
CONFigure:FOLD?
Return Parameter:
DISABLE | CVTOCC | CCTOCV
Query Example:
CONFigure:FOLD?
Return Example:
DISABLE
4.
CONFigure:FOLDT
Description:
Sets the delay time of FOLDBACK PROTECT.
Syntax:
CONFigure:FOLDT <NRf+>
Parameter:
0.01~600.00 (Unit: Sec)
Example:
CONF:FOLDT 10
Query Syntax:
CONF:FOLDT?
Return Parameter:
<NRf+>
Query Example:
CONF:FOLDT?
Return Example:
1.01
5.
CONFigure:APGVSet
Description:
Sets the action type of APG VSET.
Syntax:
CONFigure:APGVSet <Arg>
Parameter:
<Arg>: NONE | VREF10
Example:
CONFigure:APGVSet VREF10
Query Syntax:
CONFigure:APGVSet?
Return Parameter:
NONE | VREF10
Query Example:
CONFigure:APGVSet?
Return Example:
VREF10
Summary of Contents for 62000D Series
Page 1: ......
Page 2: ...Get more information by downloading Chroma ATE Solutions APP ...
Page 8: ...vi ...
Page 9: ...vii ...
Page 10: ...viii ...
Page 14: ......
Page 18: ......
Page 35: ...Overview 1 17 Figure 1 6 Dimension of 62000D 600 1200 1800 Models ...
Page 36: ......
Page 52: ......
Page 63: ...Manual Operation 3 11 Figure 3 13 Figure 3 14 ...
Page 109: ...Manual Operation 3 57 _EXT ON SIGNAL EXT ON DISABLE EXT ON ENABLE Output Output Figure 3 85 ...
Page 110: ......