Cleaning of the freeze-dryer
• Switch the freeze-dryer off by actuating the mains power switch and
disconnect the power cord from the wall outlet before cleaning.
• If the freeze-dryer has been contaminated with toxic, radioactive, or
pathogenic substances, clean the inside immediately with a suitable
decontamination agent (depending on the type of contamination, see
chapter 8.2 - "Disinfection of the drying chamber and accessories").
• Remove product residues thoroughly with a cloth.
• Open the lid/drying chamber when the freeze-dryer is not in use so
moisture can evaporate.
Pos : 174 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 175 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/080 Wartung und I nst andhaltung/ 080-0010- 0020 Eis kondensator kammer Labor @ 40\ mod_1450251422550_68.doc x @ 291807 @ 3 @ 1
8.1.1.1
Ice condenser chamber
Before each start-up, ensure that the ice condenser chamber is free from
water residues.
• Open the media drain valve to drain off any liquid. Then, close the
valve.
• If necessary, wipe the ice condenser chamber dry with a cloth.
Pos : 176 /010 Uni vers almodul e/ Seit enwechs el @ 0\ mod_1202116244312_0.doc x @ 105 @ @ 1
Pos : 177 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/080 Wartung und I nst andhaltung/ 080-0010- 0050 Belüftungs- und M edi enablauf ventil LDpl us @ 31\ mod_1408442316092_68. doc x @ 209746 @ 3 @ 1
8.1.1.2
Aeration valve, media drain valve
Contaminants such as product residues may lead to an insufficient vacuum.
In this case, the aeration valve and the media drain valve must be cleaned.
• Switch the freeze-dryer off and disconnect the mains plug.
• Remove the valve core.
• Clean the valve core and the opening with a moist cloth.
• Clean the O-rings and inspect them for any damage. Damaged O-rings
must be replaced.
1
Valve opening
2
Valve core
3
O-rings
Fig. 49: Valve opening and valve core with O-rings
(example, varies depending of the type of freeze-dryer)
• Reinsert the valve core.
• Put the freeze-dryer into operation again.
If the vacuum is still insufficient, the freeze-dryer must be checked by
qualified specialist personnel (see chapter 7.3 - "Service contact").