Freeze-dryer Beta 1-8 LSCbasic
Freeze-dryer Beta 2-8 LSCbasic
6 Operation
46 / 107
Version 03/2018, Rev. 1.5 of 09/03/2020
• sb-dc
Translation of the original operating manual
...
No
symbol
No USB storage device or LAN network connected
USB storage device connected
Active process recording on a USB storage device
Network available, but no network drive connected
Network drive connected (e.g. LyoLogplus)
Network drive connected and active process recording
Button "Schematic system diagram" (2)
The left side of the user interface shows a schematic view of the system
including all of its components. Active components are displayed in green.
Touching a component calls up its name and
→
reference designation
.
Fig. 20: Schematic system diagram with the name and reference designation of the
component
Value windows (3)
This area shows the current process data. The three value windows can be
configured as desired.
• Select the button of the value window that is to be adapted.
A dialogue box with the possible parameters opens:
– Total time (duration of the entire freeze-drying process up to this
moment)
– Section time (duration of the current section up to this moment)
– Ice condenser (temperature of the ice condenser)
– Vacuum (value of the vacuum inside the ice condenser chamber)
– Temperature
≙
vacuum (conversion of a vacuum value into a
temperature value based on the vapour pressure curve for ice and
water)