Agri-TR
65
User’s manual
07/2015
EN
Calibrations
This menu allows to modify some of the calibrations
(unapproved) and to enter new calibrations.
Click on "Meas. and calibrations" and choose "Calibrations".
Only unapproved calibrations ( "approved" box unticked) can be
modified.
The Agri-TR can not record more than 8 calibrations.
To modify a calibration, double-click on the box to be modified,
enter the modification and click on "PC => Agri-TR" to update the
Agri-TR memory.
To insert a new calibration :
- Click on "Insertion", enter the name of the new calibration and click
on "Validate". The name of the new calibration must not have more
than 2 consecutive characters in common with an approved
calibration.
- Enter the parameters of the new calibration (A0, A1, A2 and A3),
the range of the calibration (Mmin, Mmax), the HLW ref. and the
HLW. Entering Mmin,Mmax, HLW ref. and
HLW is optional.
- Click on "PC => Agri-TR" to transfer the new calibration.
Note :
For any modification on approved calibration, contact your
Chopin Technologies entitled repair organization.
Summary of Contents for Agri-TR
Page 2: ......
Page 4: ......
Page 5: ...Agri TR 5 Mode d emploi 07 2015 FR AGRI TR MODE D EMPLOI...
Page 8: ...Agri TR 8 Mode d emploi 07 2015 FR...
Page 38: ...Agri TR 38 Mode d emploi 07 2015 FR...
Page 39: ...Agri TR 39 User s manual 07 2015 EN AGRI TR USER S MANUAL...
Page 42: ...Agri TR 42 User s manual 07 2015 EN...
Page 71: ...71 Modo de Empleo 07 2015 ES AGRI TR MODO DE EMPLEO...
Page 74: ...Agri TR 74 Modo de Empleo 07 2015 ES...
Page 84: ...Agri TR 84 Modo de Empleo 07 2015 ES mediciones 2 teclas direccionales de usos m ltiples...