- 23 -
Digital Blood Pressure Monitor
MDBP13
Guidance and manufacturer’s declaration
–
electromagnetic immunity
The MDBP13 Digital Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of MDBP13 Digital Blood Pressure
Monitor should assure that it is used in such an environment.
Immunity
Test
IEC
60601
test
level
Compliance
Level
Electromagnetic environment guidance
Conducted
3 V/m
Not applicable
should be used no closer to any part of the
MDBP13
the recommended separation distance calculated
from the equation applicable to the frequency of the
RF
150 kHz
transmitter. Recommended separation distance
IEC 61000-
4-6
to
80
MHz
d
1.167
P
d
1.167
P
80 MHz to 800 MHz
d
2.333
P
800 MHz to 2.5 GHz
Radiated
RF
IEC 61000-
4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic
site survey,
a
should be less than the compliance level
in each frequency range.
b
Interference may occur in
the vicinity of equipment marked with the following
symbol:
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.