- 21-
Digital Blood Pressure Monitor
MDBP13
Declaration
Guidance and manufacturer
’s declaration – electromagnetic emission –
For all EQUIPMENT AND SYSTEMS
Guidance and manufacture
’s declaration – electromagnetic emission
The MDBP13 Digital Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer of the user of the MDBP13 Digital Blood Pressure
Monitor should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment
–guidance
RF emissions
CISPR 11
Group 1
The MDBP13 Digital Blood Pressure
Monitor use RF energy only for its internal
function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emission
CISPR 11
Class B
The MDBP13 Digital Blood Pressure
Monitor
is
suitable
for
use
in
all
establishments,
including
domestic
establishments
and
those
directly
connected to the public low- voltage power
supply network that supplies buildings used
for domestic purchases.
Harmonic emissions
IEC 61000-3-2
A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complied
Guidance and manufacturer
’s declaration – electromagnetic immunity –
For all EQUIPMENT and SYSTEMS
Guidance and manufacture
’s declaration – electromagnetic immunity
The MDBP13 Digital Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of MDBP13 Digital Blood Pressure
Monitor should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floor
are
covered
with
synthetic
material,
the
relative humidity should be
at least 30%.