ECO 6 Digital Color Doppler Ultrasound System
126
Put on medical sterile glove
Get the condom for the package.
Unfold the condom.
Load some ultrasound gel into condom.
Take the condom with one hand, and put the probe head into the condom.
Fasten the condom on the end of the probe handle.
Confirm the integrity of the condom, and repeat the above steps to the condom if any damage to the
condom is found.
9.3.3
Cleaning and Disinfecting TV and TR Probes
We strongly recommend wearing gloves when cleaning and disinfecting any endocavitary probe.
Every time before and after each exam, please clean the probe handle and disinfect the transvaginal and
transrectal probes probe using liquid chemical germicides
If the probe is contaminated with body fluids, you should disinfect the probe after cleaning.
Regard any exam waste as potentially infectious and dispose of it accordingly.
CAUTION
Since the probe is not waterproof, you should disconnect it from the system before cleaning or disinfecting.
Before and after each exam, please clean the probe handle and disinfect the transvaginal and transrectal probes
using liquid chemical germicides.
Cleaning
You can clean the transvaginal and transrectal probes to remove all coupling gel by wiping with a soft cloth
and rinsing with flowing water. Then wash the probe with mild soap in lukewarm water. Scrub the probe as
needed and use a soft cloth to remove all visible residues from the transvaginal probe surface. Rinse the probe
with enough clean potable water to remove all visible soap residues, and let the probe air dry.
CAUTION
Please remove the cover (if any) before cleaning the probe.(The cover like condom is one time usable).
When cleaning the TV and TR probes, it is important to be sure that all surfaces are thoroughly cleaned.
Disinfecting
Cidex is the only germicide that has been evaluated for compatibility with the material used to construct the
probes.
To keep the effectiveness of the disinfection solutions, a thoroughly cleaning must be done to the probe before
the disinfecting, make sure no residues remain on the probe.
Disinfecting Procedure: