![Chison ECO 6 Operation Manual Download Page 123](http://html.mh-extra.com/html/chison/eco-6/eco-6_operation-manual_2602723123.webp)
ECO 6 Digital Color Doppler Ultrasound System
110
Set as default:
set one DICOM service as default.
Send while saving:
check this item and enable DICOM storage while saving image or cine, send clip or
image according to activated function.
Send after ending exam:
check this item and enable DICOM storage while saving image or cine, send clip
or image according to activated function.
7.7.1
Add/Edit DICOM Function
Fig. 46 Add DICOM Setting Interface
Press
[Add]
button, then system pops up the DICOM Setting interface.
Default:
check this option, set DICOM function as default.
Service:
choose DICOM Storage, DICOM Worklist, DICOM Print or DICOM SR.
Server AE Title:
input DICOM server AE name.
Server IP address:
input DICOM IP address.
SCP port:
input DICOM server SCP port.
Timeout:
set the delay time of DICOM..
Compression Mode:
check this option, set Uncompressed or JPEG according to the corresponding
requirements.
Compression Ratio:
after selecting the JPEG as the compression mode, choose the appropriate
compression ratio.
Max Framerate:
check this option, select proper feamerate according to the requirement.
Tip:
choosing the DICOM Print type must be before relative print setting.
7.8
NET Work
Set the unit’s and target unit’s IP and do the connection testing. And network storage settings, details see the
appendix: Procedures of set network sharing in ECO series.