ECO Series VET Digital Color Doppler Ultrasound System
38
Baseline
: adjust the baseline to accommodate faster or slower blood flows to eliminate aliasing.
Angle
: optimize the accuracy of the flow velocity. It estimates the flow velocity in a direction at an angle to
the Doppler vector by computing the angle between the Doppler vector and the flow to be measured. This is
special useful in vascular applications where you need to measure velocity.
Doppler Gain
: allow you to control the background information of spectral.
Sweep Speed
: control speed of spectral update.
Doppler Sample Volume Gate Position and Size (Trackball and ENTER)
Move the sample volume on the B mode’s Doppler cursor. The gate is positioned over a specific position
within the vessel.
To move Doppler cursor position, turn the trackball left or right until positioned over the vessel.
To move sample volume gate position, move the trackball up or down until positioned inside the vessel.
To size sample volume gate, press
[ENTER]
key to toggle trackball function from sample volume gate
positioning to sizing, then move the trackball to change sample volume gate size.
5.6
B Image Adjustment
5.6.1
Frequency
In real status, press the first line key of [
Frequency
] to increase the frequency, and press the second line key of
[
Frequency
] to decrease the frequency.
5.6.2
Dynamic
Dynamic range is used for adjusting the contrast resolution of B mode image and mode image, compressing or
enlarging the display range of gray scale.
At the real-time status, press the first line key of [
Dynamic
] to increase the Dynamic, and press the second line
key of [
Dynamic
] to decrease the Dynamic. The range is 30 to 90.
5.6.3
i-Image
In real status, press the corresponding key of [
i-Image
] to adjust, the range is 0~3.
5.6.4
Compound
In real status, press the corresponding key of [
Compound
] to adjust, and the range is 0~3.
The SRA can’t be edited after opening the compound.
5.6.5
SRA
In real status, press the corresponding key of [
SRA
] to turn on or off.
5.6.6
M Speed Adjustment