ECO Series VET Digital Color Doppler Ultrasound System
148
Appendix B: Guidance and Manufacturer’s Declaration
1. Guidance and manufacturer’s declaration – electromagnetic emissions
The ECO Series VET is intended for use in the electromagnetic environment specified below. The
customer or the user of the ECO Series VET should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic
environment
–
guidance
RF emissions
CISPR 11
Group 1
The ECO Series VET uses RF
energy
only
for
its
internal
function.
Therefore,
its
RF
emissions are very low and are not
likely to cause any interference in
nearby electronic
equipment.
RF emissions
CISPR 11
Class A
The ECO Series VET is suitable for
use in medical establishments
including domestic establishments
and those directly connected to the
public high-voltage power supply
network that used for non-domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
2. Guidance and manufacturer’s declaration – electromagnetic immunity
The ECO Series VET is intended for use in the electromagnetic environment the ECO Series VET
should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic
environment –
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic
material,
the
relative humidity should
be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
±2 kV for power
supply lines
±1 kV for input/output
lines
Mains
power
quality
should be that of a typical
commercial or hospital
environment.