Combination 860
33
Width:
pressing this button opens a new window in which the pulse width
may be set from 20 to 400 µs in 5 µs increments by pressing the numerical
digits and then pressing OK.
Frequency:
pressing this button opens a new window.
The pulse frequency may be set from 1 to 250 Hz in 1 Hz increments by
pressing the numerical digits and then pressing OK.
Mode control:
The TENS output may be continuous, burst or surged. Each
modulation type is selected by pressing the Mode button.
CC/CV:
The output may be set to be constant current (CC) or constant
voltage (CV) by pressing the CC/CV button.