Combination 860
31
Russian set-up
Treatment time:
Is selected in the same way as previously.
Burst:
The burst frequency may be set from 1 Hz to 100Hz by pressing the
Burst button. 0-10 Hz is in 1 Hz steps, 10 – 50 Hz in 5 Hz steps, and 50 –
100 Hz in 10 Hz steps.
Ratio:
The medium frequency bursts used for russian stimulation are
surged to produce work and rest periods. The surge time (work) is fixed at
10 seconds. The ratio sets the off or rest time in ratio to the work period.
For example, if the ratio is set to 1:4 then the work period is 10 seconds
and the rest period is 40 seconds. Press the ratio button to set the ratio to
any integer value between 1:1 and 1:5.
When running, the burst status is displayed in the top right corner of the
display -
CC/CV:
The output may be set to be constant current (CC) or constant
voltage (CV) by pressing the CC/CV button. In constant current mode the
electrode impedance is monitored and if the impedance for either channel
is too high then the output is terminated and an alarm sounded. In constant
voltage mode, if the electrode impedance rises then the output current is
automatically reduced.