5.7.2 Digital input data
R/W
……
R:READ
,
W:WRITE
Reference
No.
Applicable
function code
R/W Description
Details
10009
10010
10011
10012
02 R
Remote contacts 1 Status
Remote contacts 2 Status
Remote contacts 3 Status
Remote contacts 4 Status
Status of remote contacts input
0 : OFF
1 : ON
10101
10102
02
R
CH1 Status 1
Status expression with 2 BIT
00 : Measured value
01 : Operation data
Error code : 01H, 02H, 03H
10105
10106
10107
10108
02
R
CH1 Status 2
Status expression with 4 BIT
0000
:
Normal data
0001
:
+ Over range
0010
:
- Over range
0100
:
Burnout
1000 : Invalid data (initialization, during data
logging, no range programming)
Error code : 01H, 02H, 03H
10109
10110
10111
10112
02 R
CH1
Alarm Level 1
CH1
Alarm Level 2
CH1
Alarm Level 3
CH1
Alarm Level 4
Status
0 : No alarm
1 : Alarm on
Error code : 01H, 02H, 03H
10117
10118
02
R
CH1 Status 1
Status expression with 2 BIT
00 : Measured value
01 : Operation data
10 : Communications input data
Error code : 01H, 02H, 03H
10121
10122
10123
10124
02
R
CH1 Status 2
Status expression with 4 BIT
0000 : Normal data
0001 : + Over range
0010 : - Over range
0011 : Burnout
1000 : Invalid data (initialization, during data logging,
no range programming)
Error code : 01H,02H,03H
10125
10126
10127
10128
02 R
CH2
Alarm Level 1
CH2
Alarm Level 2
CH2
Alarm Level 3
CH2
Alarm Level 4
Status
0 : No alarm
1 : Alarm on