INST.No.INE-508-P0
-6-
3.4 Layout of connecting cables
For connecting the connecting cables, be careful of the following points.
3.5 Connections
(1) Connections of terminal connection type
By removing the terminal board cover on the
upper side of the detector unit, 5 pieces of
terminals are arranged as shown in the right
figure.
Connect wires of the connecting cable to each
terminal.
Since the wires of the connecting cable are
identified by colors, connect them as shown in
the right figure.
(2) Connection of connector connection type
The connection is completed by coupling the connector of the connecting cable to the
connector on the upper side of the detector unit.
x
For the coupling, position key slits of both connectors and push the cable connector,
and turn the detachable ring clockwise until it is locked.
x
For decoupling, turn the detachable ring counterclockwise for unlocking and pull the
cable connector.
3.6 Operation
Since the drive power is supplied to the detector unit from the relay box, when the relay box is
turned on, the detector unit automatically becomes ready to measure.
However, when the protective case and/or the air-purge hood are used, confirm that cooling
water and/or purge air are supplied in advance.
By starting up the software of the high-speed display/processing unit and then clicking the
measurement start button, the detector unit is put into operation.
The power for the detector unit
is supplied from the relay box
through the connecting cable.
The terminal board is exclusively
used for connection with the
relay box. Do not connect a
power line, etc.
3. Detector unit
(1) Separate the cables from induction heating equipment and power lines.
(2) Keep the cables free of deposit of water, oil, etc.
(3) Don’t bend the cables extremely or apply any excessive force to them.
(4) Avoid laying the cables at a place in high temperature more than 70°C.
(5) For laying the cable permanently, protect them with conduits, etc.
The outside diameter of the connector attached to the cables is about
ø23mm.
Caution
Caution
R e m a r k s
Black
White
Red Yellow Shield
+
-
AC100V
OUTPUT POWER EARTH