CHINESPORT Spa – ITALIA
AV3
42/47
AV3 - User Manual rev.1.2 - EN.docx
We suggest to place the labels in a comfortable place in order to consult the
charging instructions in an easy way.
CHARGING INSTRUCTIONS
1)
SET THE WHITE POWER BUTTON ON “I”
2)
CONNECT THE BATTERY CHARGER TO THE POWER SUPPLY AND
TO THE CHARGING SOCKET
FOR FUTHER INFORMATION PLEASE REFER TO THE USER MANUAL
DAILY RECHARGING IS RECOMMENDED
CHARGING INSTRUCTIONS
1) SET THE WHITE POWER BUTTON ON “I”
2) CONNECT THE BATTERY CHARGER TO THE POWER SUPPLY AND TO THE
CHARGING SOCKET
FOR FUTHER INFORMATION PLEASE REFER TO THE USER MANUAL
DAILY RECHARGING IS RECOMMENDED
ELECTROMAGNETIC COMPATIBILITY INFORMATIONS
The manufacturer declares that AV3 including his original accessories supplied with
the device complies with the requirements of EN 60601-1-2 for electromagnetic
compatibility. The battery charger is only for charging battery when device is not in
use, and has its own certification.
10.1
Manufacturer declaration – Electromagnetic emissions
AV3 is intended for use both in the home and in the clinic, in an electromagnetic
environment described in the following table. The user of the AV3 must guarantee
its use in the appropriate environment.
Emission test
Conformity
Recommended electromagnetic environment
Irradiated / Conducted RF
emissions CISPR 11
Group 1
AV3 uses RF energy only for its internal
functions. Therefore, its RF emissions are very
low and are unlikely to cause some
interference in nearby electronics.
Irradiated / Conducted RF
emissions CISPR 11
Class B
AV3 is suitable for use in all environments
including home environments and all those
directly connected to the low voltage power
supply network that supplies home-based
buildings.
Harmonic emissions
IEC61000-3-2
Not Applicable
Voltage fluctuations and
flickering IEC61000-3-3
Not Applicable
10.2
Manufacturer declaration – Electromagnetic immunity
AV3 is intended for use both in the home and in the clinic, in an electromagnetic
environment described in the following table. The user of the AV3 must guarantee
its use in the appropriate environment.
Immunity test
CEI 60601-1-2 test levels
Electromagnetic environment -
Recommendations
Electrostatic
discharges (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Floors should be covered with wood,
cement or ceramic tiles. If the floors are