At this point the unit will wait to receive the starting and ending pages to erase. Page
numbers are 512 bytes and can be calculated by dividing the starting page memory location
by 512. The following table will help you determine the table page numbers.
Table
Bytes
/line
Address
Start
Page
End Page
Erase Command
Sensor Labels
8
6800
34
37
*3438
Groups
1
7000
38
3B
*383C
Label Positions
1
7000
38
3B
*383C
Alarms
1
7800
3C
3F
*3C40
Channels
1
7800
3C
3F
*3C40
Options
1
7800
3C
3F
*3C40
Lookup Table 0
8
0800
04
07
*0408
Lookup Table 1
8
1000
08
07
*080C
Lookup Table 2
8
1800
0C
0B
*0C10
Lookup Table 3
8
2000
10
13
*1014
Lookup Table 4
8
2800
14
17
*1418
Lookup Table 5
8
3000
18
1B
*181C
Lookup Table 6
8
3800
1C
1F
*1C20
Lookup Table 7
8
4000
20
23
*2024
Lookup Table 8
8
4800
24
27
*2428
Lookup Table 9
8
5000
28
2B
*282C
Lookup Table 10
8
5800
2C
2F
*2C30
Lookup Table 11
8
6000
30
33
*3034
To erase the memory you
must enter a “*” symbol followed by the start and end page values
with no spaces. For example to erase the Alarms table enter:
•
*3C40
To erase Lookup Table 3 enter:
•
*1014
Once you have entered the 5 characters the unit will erase the memory and respond with
•
ERASED->
At this point the unit will wait to receive the data to reprogram the unit. Use the
HyperTerminal Send Text File function to transfer the appropriate file to reprogram the
memory. As the file is transferring data the unit will display the memory address for each