Chemtrac HydroACT Quick Start Manual Download Page 11

11 

 

Calibration Procedure 

1.

 

When a new membrane cap is installed, it may take up to 30 minutes before the chlorine probe starts to 

register a reading (>0 mg/l). Although the after 2 hours the reading should be mostly stabilized (80-

90%),it is recommended to allow the chlorine probe with new membrane cap to operate for a minimum 

of 6 hours before performing an initial calibration.    

 

2.

 

Collect a sample at the chlorine probe chamber and test for residual chlorine level with an approved 

benchtop or handheld device (e.g. DPD test kit).  It is important to test a sample taken at the same 

location.  Write down the test result and immediately proceed to step 3.   Note: For greatest accuracy, 

consider performing 3 tests and ensure acceptable repeatability, and then average the results.   

 

3.

 

Calibration is initiated from the Sensor Overview screen which is accessed by pressing Menu and then 
selecting the Free or Total chlorine tile.      
 

 

 

4.

 

Press the keypad button directly below the “Calibrate” icon on the display.  Then select “Calibrate Span.”   
Follow the user prompts.    
 

 

     

 

 

5.

 

When prompted to “Enter Calibration Value”, press the center (Select) button to enter Edit mode.  This 
will cause the text to go from blue to purple and the Home, Options, and Next buttons will turn gray as 
seen above.   Once in Edit mode, use the arrows on the keypad to adjust the reading to match the DPD 
result.  Press the center (Select) button again to exit the Edit mode, and then press Next.    A prompt will 
appear asking if the electrolyte and membrane were replaced.  Answer the prompts appropriately. 
 

6.

 

This completes the calibration.   The calibration will need to be verified on a routine basis and should not 
exceed 30 days.   Local regulations regarding measurement and reporting of residual chlorine levels may 
require daily to weekly calibration verification.  Consult local agencies (e.g. EPA, DEQ, Health & Safety) for 
more info.  

 

 

 

Summary of Contents for HydroACT

Page 1: ...Quick Start Guide HydroACT Analyzer with Free or Total Chlorine Sensor Version 03 10 2020...

Page 2: ...2...

Page 3: ...This Quick Start Guide is a supplement to the User Manual and is intended to give an overview of the product accessories and set up prior to installation The user is advised to review the full produc...

Page 4: ...easurement accuracy With the vented design it is very important to not have any backpressure on the drain line as this would limit how much flow would be able to pass through the probe chamber before...

Page 5: ...n Line Connections Flow Cell with Single Probe Chamber HydroACT Analyzer A mounting template and screws are provided with analyzer Mounting Holes 1 7 6 in 194 mm 2 9 4 in 238 mm Dimensions 1 4 1 in 10...

Page 6: ...ce is applied In lifting the cover position it to insert the display posts into the post holes as indicated allowing the cover to temporarily rest in the raised position After completing wiring inside...

Page 7: ...shows where the connections are located for the 4 20 mA and 4 20 mA on output card See next page for the analog output setup and scaling instructions Relays and Digital Inputs Optional Relays and Digi...

Page 8: ...r or analog output was supplied the user should verify the settings before wiring User can change the mapping and output scaling if desired Note Press the Next button to navigate through the 3 setup s...

Page 9: ...with a Total Chlorine probe should be labeled Total or have the part number ECP1 4 Damage can occur to the probe if the incorrect electrolyte is used Position the probe vertically so that the electro...

Page 10: ...is typically adequate to keep air bubbles from forming on the membrane Sample Flow Considerations For proper operation a minimum of 250 ml min sample flow is required Large changes in flow e g from 25...

Page 11: ...hich is accessed by pressing Menu and then selecting the Free or Total chlorine tile 4 Press the keypad button directly below the Calibrate icon on the display Then select Calibrate Span Follow the us...

Page 12: ...e tip where the white dot is located See page 10 of this document for instructions on how to position probe and adjust flow to deal with air bubbles 5 Ensure the cable is screwed all the way onto the...

Reviews: