Chemtrac HydroACT Quick Start Manual Download Page 10

10 

 

Proper Positioning of Probe into the Flow Cell Probe Chamber 

Once the probe is assembled, insert the probe through the compression fitting and 
tighten  the fitting  around the  probe  so it  is  snug  but  still  allows  the  probe to slide 
through when pressure is applied.     

Lower the probe into the flow chamber and slide the probe through the fitting until the 
tip  of  the  probe  tip  is  aligned  with  the  small  horizontal  tube  that  connects  to  the 
adjacent chamber (see pictures below).  This position allows the water that is exiting 
this tube to flow across the surface of the membrane which keeps air bubbles from 
forming on the membrane tip (where the white dot is located). This is where chlorine 
passes through the membrane.   Readings can become unstable and drop off towards 
zero if bubbles form on the membrane cap tip (white dot).   If bubbles are seen forming 
on the tip, ensure the probe is properly positioned and increase the flow until bubbles 
dislodge.  With the probe properly positioned, a flow rate of 500 to 750 mL/min is 
typically adequate to keep air bubbles from forming on the membrane.    

 

 

 

  

  

  

 

 

Sample Flow Considerations 

For proper operation, a minimum of 250 ml/min sample flow is required.  Large changes in flow (e.g., from 250 to 
1000 ml/min) can impact the reading.  If flow exceeds 1000 ml/min, the probe chamber may overflow.  If overflow 
cannot be tolerated, or if sample flow is too variable due to changes in line pressure, then a pressure regulator, 
constant head pressure device, or Dole valve (flow regulator) is recommended.    

Drain Line Considerations 

Sample flow through the probe chamber will be restricted if there is any backpressure on the drain line.  To avoid 
backpressure, make sure to keep drain lines short and as vertical as possible, and make sure the end of the drain 
tubing is not submerged under water.   To test for proper draining, set the flow for 1000 ml/min. If the probe 
chamber overflows at this flow rate, then there is backpressure on the drain line.  This can be verified by unhooking 
the drain tubing from the flow cell.  If the flow cell stops overflowing when the drain line is disconnected, then the 
drain line has too much back pressure.   

Note:  An air bubble covering this white dot will cause the chlorine reading to drop off.    

Summary of Contents for HydroACT

Page 1: ...Quick Start Guide HydroACT Analyzer with Free or Total Chlorine Sensor Version 03 10 2020...

Page 2: ...2...

Page 3: ...This Quick Start Guide is a supplement to the User Manual and is intended to give an overview of the product accessories and set up prior to installation The user is advised to review the full produc...

Page 4: ...easurement accuracy With the vented design it is very important to not have any backpressure on the drain line as this would limit how much flow would be able to pass through the probe chamber before...

Page 5: ...n Line Connections Flow Cell with Single Probe Chamber HydroACT Analyzer A mounting template and screws are provided with analyzer Mounting Holes 1 7 6 in 194 mm 2 9 4 in 238 mm Dimensions 1 4 1 in 10...

Page 6: ...ce is applied In lifting the cover position it to insert the display posts into the post holes as indicated allowing the cover to temporarily rest in the raised position After completing wiring inside...

Page 7: ...shows where the connections are located for the 4 20 mA and 4 20 mA on output card See next page for the analog output setup and scaling instructions Relays and Digital Inputs Optional Relays and Digi...

Page 8: ...r or analog output was supplied the user should verify the settings before wiring User can change the mapping and output scaling if desired Note Press the Next button to navigate through the 3 setup s...

Page 9: ...with a Total Chlorine probe should be labeled Total or have the part number ECP1 4 Damage can occur to the probe if the incorrect electrolyte is used Position the probe vertically so that the electro...

Page 10: ...is typically adequate to keep air bubbles from forming on the membrane Sample Flow Considerations For proper operation a minimum of 250 ml min sample flow is required Large changes in flow e g from 25...

Page 11: ...hich is accessed by pressing Menu and then selecting the Free or Total chlorine tile 4 Press the keypad button directly below the Calibrate icon on the display Then select Calibrate Span Follow the us...

Page 12: ...e tip where the white dot is located See page 10 of this document for instructions on how to position probe and adjust flow to deal with air bubbles 5 Ensure the cable is screwed all the way onto the...

Reviews: