Battery Replacement
To ensure that the instrument is waterproof:
• seal ring (A) must be in position
• battery compartment cover (B) must be fixed with the four screws
To Set Zero
1. Press the ON/OFF key.
2. The display will show “
O3
”.
3. Insert the ZERO ampoule (supplied in Vacu-vials
®
kit), flat end first, into the
sample cell compartment (with mild downward pressure), making sure that it is
fully seated.
4. Place the light shield over the ZERO ampoule.
5. Press the Zero/Test key. The “
O3
” symbol will flash for approximately 8 sec-
onds, then the display will show “
SEt
”.
To Generate Reagent Blank Value
1. Insert an
unsnapped ozone Vacu-vial ampoule
, flat end first, into the sample
cell compartment (with mild downward pressure), making sure that it is fully
seated.
2. Place the light shield over the ampoule.
3. Press the Zero/Test key. The “
O3
” symbol will flash for approximately 8 sec-
onds, then the display will show “
tESt
“.
Note:
The instrument has stored a reagent blank value for this ampoule.
To Obtain Test Result
1. Follow the Test Procedure in the Ozone Vacu-vials test kit (Cat. # K-7433) using
the ampoule from which the reagent blank value was generated above.
2. Insert the resulting Ozone Vacu-vial ampoule, flat end first, into the sample cell
compartment (with mild downward pressure), making sure that it is fully seated.
3. Place the light shield over the test ampoule.
4. Press the Zero/Test key. The “
O3
” symbol will flash for approximately 3 sec-
onds, then the sample test result will appear in the display as ppm (mg/Liter).
5. To perform the next test, remove the test ampoule from the sample cell com-
partment. Press the Zero/Test key. The display will show “
SEt
”. Return to
To
Generate Reagent Blank Value
.
Operating Tips
• Upon startup, the photometer automatically proceeds to the zeroing process.
Every time the photometer powers on, it must be re-zeroed.
• To re-zero the photometer, it must be turned off and back on again.
• A series of readings can be taken without re-zeroing, as long as the photome-
ter stays on during the series.
• Protect photometer from extreme humidity, corrosive fumes and dusty areas.
Store in a cool, dry place.
• Remove the batteries when photometer is not in use.
• Press the
!
key to turn the display back light on or off.
• When moving the photometer from one temperature extreme to another, wait
at least 10 minutes before use to allow photometer to come to temperature
equilibrium.
• Contamination of the optics in the sample chamber will result in incorrect meas-
urements. The windows in the sample chamber should be checked at regular
intervals and cleaned as necessary. Use a soft moist cloth or cotton swab for
cleaning purposes.
• If the sample cell adapter has been removed, it must be replaced with proper
orientation, aligning the triangle on the adapter with the triangle on the pho-
tometer.
Displays and Troubleshooting
E01:
Light absorption too great (dirty optics)
E20/E21:
Too much light reaching detector
E22 or Battery Icon:
Battery should be replaced
E27/E28/E29:
Instrument zeroed incorrectly, misaligned adapter, vial not
properly seated, dirty optics or failing light source.
Hi/E03:
Measuring range exceeded or excessive turbidity
Lo:
Test result has a negative value (less than 0 ppm)
Specifications
Auto Shutoff:
After 15 minutes of non-use
Optics:
610 nm LED/interference filter and photosensor in transparent sample
chamber
Operating Temp.:
5 to 40
°
C (41 to 104
°
F)
Battery:
4 AAA batteries (approx. 5,000 tests or 17 hours)
Waterproof:
Floating, IP68 (1 hour at 0.1 meter)
Wavelength Accuracy:
± 1 nm
Photometric Accuracy*:
3% full scale (T = 20 - 25° C / 68 - 77° F)
Photometric Resolution:
0.01 A
Ambient Conditions:
Temperature 5 - 40° C / 41 - 104° F
Rel. humidity 30 - 90 % (non-condensing)
CE:
Certificate of Declaration of CE-Conformity available upon request.
* Measured with Standard Solutions
Instrument
Back
Batteries
Notch
Screws
Seal Ring (A)
Battery
Compartment
Cover (B)