6.7. ‘Advanced’ Page
The advanced page contains functions that will change the how the flightDAQ acts and how it
applies various calculations to its data, this page should only be used by users who fully understand
what they are changing.
6.7.1. Filtering
The flightDAQ has settings to allow the user flexible control over the data throughput of the device.
Averaging options allow the preference of noise reduction over time domain response, the frequency
of calibration temperature compensation renewal may be chosen as may be the size of the average
(and hence time taken) for the rezero routine. Table 5.9 details the function of the signal parameter
option controls.
Figure 5.9, Filtering group
Control
Function
'Pressure Input Impulse Filter'
Check box to apply impulse filter to pre-calibration data - will remove
single impulse noise events in the pressure data.
'Temperature Input Impulse Filter'
As above but with temperature data.
'Pressure input average samples'
Selects the number of samples for a moving average of pre-calibration
pressure data.
'Temperature input average samples'
As above but with temperature data.
'Temperature Compensation'
Selects the temperature compensation scheme for the calibration.
'Continuous' repeatedly rebuilds the calibration data on a channel by
channel basis without interrupting the flow of data. 'With zero only'
rebuilds the table when a user rezero is issued, after the rezero has
been actioned.
‘Apply’
Applies the settings to the local settings memory
Table 5.8, Filtering group settings