– 5 –
2.4 GAIN Key
The TI-007X has an adjustable gain feature for additional versatility to control
the overall output of the gauge, to compensate for various materials and
applications. There are five gain settings (VLOW, LOW, MED, HIGH, VHI)
when pulse-echo or thin PLAS mode is selected, and three gain settings (LOW,
MED, HIGH) for all other modes where automatic gain control is used. Refer to
8.1 for details regarding the gain feature..
2.5 SCAN Key
Th offers a high speed scan feature. This feature allows for scanning larger areas
on a given test material, while still offering reasonable representation of
thickness.
2.6 +/- Increment/Decrement Key’s
The +/- Keys are used to increment/decrement values,
navigate menus, and select menu options.
2.7 MODE Key
The MODE key toggles the measurement modes. If a delay line transducer (dLY)
is selected from the PROB menu option, echo-echo (E-E) and interface-echo (I-E)
modes will be toggled. If a contact transducer (Con) is selected, pulse-echo (P-E) and
echo-echo (E-E) modes will be toggled. Finally, if the plastic transducer (PLAS) option
is selected, the thickness options thin (thin) and thick (thK) will be toggled. The
measurement modes and transducer styles offer the flexibility to address specific
application scenarios. Refer to section 7.0 for an explanation on the various calibration
options.
2.8 MENU Key
The MENU key is used to access and set all of the additional features of the
TI-007X that are not at the top level of the keypad with a dedicated key. The
features and setting are outlined in the table below: