–32 –
5. When the desired VX setting is displayed, press the
key to set the status
and return to the measurement screen.
8.10 Lock
The lock feature was built into the TI-007X for the purpose of locking the operators
out of editing any of the gauge settings, for purposes of consistency between operators.
When the lock feature is enabled, the gauge calibration functionality cannot be altered,
as well as any of the individual features in the gauge. The only keys that are always
unlocked are the power and probe zero keys, as these must remain unlocked for
measurement functionality.
The procedure to enable/disable the lock feature is outlined below:
1. With the TI-007X powered off, press and hold down the
key while
powering the TI-007X on . The lock icon will
be
illuminated on the display.
2. To unlock the TI-007X repeat step one, but hold down the
key while
powering the TI-007X on.
8.12 Factory Defaults
The TI-007X can be reset to factory defaults at any time to restore the original gauge
settings. This should only be used if the gauge is not functioning properly, or perhaps
multiple features have been enabled and a clean start is needed.
The procedure to reset the gauge is outlined below:
1. With the TI-007X powered off, press and hold down the
and
keys
while powering the TI-007X on .
NOTE:
Once the measurement screen is displayed the keys can be released.