5

4.0

QUICK START INSTRUCTIONS

NOTE: 

If tension meter is equipped with a Thickness

Compensator, first set up Compensator as indicated in
3.0.

1. Prior to engaging the material under test, part the 

3-roller system by pushing and holding the Thumb 
Piece forward.

2. Keeping the outer rollers extended, bring the

instrument behind or under the filament and move 
it so that the Filament Guide bars contact the 
process material.  

3. Slowly release pressure on the Thumb Piece until 

the outer rollers return to their original position. 
This will automatically direct the material into the 
guide roller grooves.

4. The scale pointer will show line tension directly in 

grams or kilograms.

5. To remove the instrument from the process 

material, push the Thumb Piece forward again to 
open the 3-roller system. With the outer rollers
extended, move the instrument away from the 
material.  

6. Remove the instrument and SLOWLY release

pressure on the Thumb Piece, allowing the outer 
rollers to return to their original position.

CAUTION: Do not let the Thumb Piece snap 
back as this may affect calibration and also 
damage the instrument.

8

5.2  Special Calibration

Special Calibration is optionally available for a specific
material type and size so long as the material fits the
mechanical limits and range of the instrument. The
customer must supply a 10-ft sample (3 meters) of the
material.  Such calibration can be performed with or
without using the Thickness Compensator if one has
been provided with the instrument. If not specified, 
the calibration facility will use its best judgment.   

Note: 

Special calibration is available for a single 

sample only.  

5.3  NIST Calibration

While all DXN instruments are furnished with a
Calibration Certificate which certifies that they have
been calibrated in accordance with factory procedures
and were found to meet all published accuracy specifi-
cations, such calibrations do not fulfill ISO-9000
requirements since no record of measured values are
kept or are submitted. Where ISO-9000 requirements
are to be met, NIST calibration is optionally available
but must be specified at time of order placement or
after repair.  

Summary of Contents for DXN

Page 1: ...HECK LINE BY ELECTROMATIC DXN Operating Instructions Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail info abqindustrial net ...

Page 2: ...racy 1 of full scale or 1 graduation mark on scale Diameter of Scale 54 mm Maximum Speed standard 2000 m min optional 3500 m min K optional 5000 m min HS Outer roller distance c c DXN 30 to DXN 1000 EX 38 mm DXN 200 to DXN 10K 100 mm DXN 2K 150 or 200 mm DXN 30K to DXN 50K 250 mm Roller material standard Hard coated aluminum optional Hardened steel ST Ceramic CE Phenolic PH Housing material Die ca...

Page 3: ...3 Air Dashpot Damping DXN Tension Meters incorporate a moderate amount of movement damping to smooth out some high fre quency tension fluctuation However where process fluctuations result in rapid pointer swings which make it difficult or near impossible to read the scale an optionally available Air Dashpot Damping mechanism can be provided to slow down the pointer swings to permit a reading In se...

Page 4: ...n high range instruments the return spring force must be strong enough to bring the outer rollers back to their original position during measurement The average person will find it difficult to operate the mechanism just with thumb pressure on the Thumb Piece alone This is especially true with models rated 20 Kg and higher Adding the Lever Assembly shown here will greatly reduce the effort require...

Page 5: ...e 5 4 Calibration Verification Schedule Frequency of calibration accuracy verification depends on many factors These include frequency and extent of tension overloads operating speeds length of oper ating times environment care in handling etc Such determination is best made by the user s Quality Assurance Department based on the user s experience However a quick calibration check near the anticip...

Page 6: ...to return to their original position CAUTION Do not let the Thumb Piece snap back as this may affect calibration and also damage the instrument 8 5 2 Special Calibration Special Calibration is optionally available for a specific material type and size so long as the material fits the mechanical limits and range of the instrument The customer must supply a 10 ft sample 3 meters of the material Such...

Page 7: ...quest 5 1 Calibration with Factory Standard Filaments All DXN Tension Meters are calibrated with laboratory weights suspended from factory Standard nylon monofilament sizes as noted in the table below Any difference in process material size and rigidity from the Standard materials may result in some error For over 95 of applications the error is not significant and can be ignored since readings ar...

Reviews: