Electromatic Equipment Co., Inc. 

-4- 

7.   Fill  up  the  measuring  device  with  the  sample  pellets. 

The  filling  needs  to  be  done  slowly  and  constantly  to 
ensure reproducible results. 

 
 
 
 
 
 
 
 
 
8.   The display shows the measuring result. 
 
 
 
 
 
9.   To save the results in the storage menu press the  

  

(

  button).  The  storage  was  successful  when  the 

number  in  front  of  the  symbol 

  increased.  To  reach 

the  storage  menu  please  press  (

)  until  the 

 

appears. 

 
10.   To name the saved results press the 

 button.  

 
11.   Empty  the  humimeter  and  ensure  that  no  pellets  rests 

are accumulated in the measuring chamber. 

 
 
 

Measuring range

If  the  measure  value  is  blinking,  the  valid  measuring  range 
is  exceeded  (limits  see  list  on  page  5).  In  this  case  the 
accuracy will be decreasing. 
 
 
 
 
 
 
 
 
 
 

 

Electromatic Equipment Co., Inc. 

-13- 

11.0  MOST COMMON REASONS FOR MISS READINGS 

 
 

 

Product temperature out of application range 

 

Material  below  0°C  resp.  above  40°C  (32  to  104  °F)  may  cause  faulty 
measurements.  The  storage  of  cold  material  in  a  warm  storage  area  usually 
creates condensed water which may lead to major measuring errors. 

 

 

Not adjusted pellets 

 

Let  your  humimeter  BP1  adjust  to  the  surrounding  temperature  of  the  material 
for approximately half an hour. 

 

A  very  high  temperature  difference  has  a  negative  effect  on  the  stability  of  the 
measurement result. 

 

 

Wrong calibration curve 

 

Before  you  measure  your  sample,  double  check  the  correct  selection  of  the 
calibration curve. 

 

 

Wrong filling quantity 

 

Fill in exactly 280 gram (± 1.0 g) of wood pellets in the measuring chamber. 

 

 

Wet or mouldy material leads to wrong values!! 

 

 

Frozen measuring material leads to wrong values!! 

 
 
 
12.0  DEVICE MAINTENANCE INSTRUCTIONS 

 
 
To provide a long lives of your device please do not expose it to strong mechanical 
loads or heat e.g. dropping it or direct sunlight exposure. 
 
Clean your device using a 

dry cloth. 

The measuring chamber needs to be cleaned 

with a 

dry and soft brush

 
Any kind of wet cleaning damages the device. The instrument is not rainproof. Keep 
it in dry areas. 
 
We recommend an interval ISO-conforming device check with the drying method. 
 
Electromatic also provides a test with a calibration certificate for a fee. 

Summary of Contents for BP1

Page 1: ...cedure 3 3 0 Calibration curves 5 4 0 Menu level overview 6 5 0 Keypad symbols 7 6 0 Changing batteries 8 7 0 Determination of the material moisture 8 8 0 Transfer saved data to the PC 9 9 0 Print saved data 11 10 0 Technical data 12 11 0 Most common reasons for miss readings 13 12 0 Device maintenance instructions 13 13 0 Warranty 14 ...

Page 2: ...nc 2 1 0 DESIGN OF THE DEVICE Batteries slot flipside Measuring chamber Optional PC interface Name of calibration Moisture content POWER button Filling assistance Change the calibration curve Electromatic Equipment Co Inc 15 NOTES ...

Page 3: ...s do not allow the exclusion or limitation of incidental or consequential damages so the above limitation may not apply to you The duration of any implied warranty including without limitation fitness for any particular purpose and merchantability with respect to this product is limited to the duration of the foregoing warranty Some states do not allow limitations on how long an implied warranty l...

Page 4: ...rage area usually creates condensed water which may lead to major measuring errors Not adjusted pellets Let your humimeter BP1 adjust to the surrounding temperature of the material for approximately half an hour A very high temperature difference has a negative effect on the stability of the measurement result Wrong calibration curve Before you measure your sample double check the correct selectio...

Page 5: ...ring range limits 280g wood pellet 280g 3 up to 20 250g spec pell 1 250g 3 up to 20 250g spec pell 2 250g 3 up to 20 250g spec pell 3 250g 3 up to 20 reference Explanation of the curves 280g wood pellet To analyze the moisture of Ø6 and Ø8 mm wood pellets with a maximum length of 30mm According to norm CEN TS 14961 If you do not want to measure wood pellets or wood pellets with smaller bulk densit...

Page 6: ...rial interface module in combination with Schaller thermo printer To print your saved logs connect the printer to the humimeter serial interface using the cable that was delivered with your device To switch on the printer press the ON OFF button If the printer is correctly switched on the green LED blinks regularly If it does not replace the batteries and try again Press the key until you reach th...

Page 7: ...ow the port number COM1 COMX You can find this setting in the device manager directory ports COM LTP See print screen below The shown port number in the device manager COMx in our example COM6 for the USBRS232 adapter has to be selected in the software humimeter LogMemorizer directory Extras Settings Device Close the dialog using the button OK Electromatic Equipment Co Inc 7 5 0 KEYPAD SYMBOLS Mea...

Page 8: ...ring results Each device includes a serial number and the guarantee stamp If those are broken no claims for guarantee can be made In case of a faulty device please contact Electromatic F x 100 Mn Mt Mn Electromatic Equipment Co Inc 9 8 0 TRANSFER SAVED DATA TO THE PC Only possible with humimeter serial interface module To send your saved logs to your PC connect the humimeter device to your PC usin...

Page 9: ...BP1 PELLET MOISTURE METER OPERATING INSTRUCTIONS Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail info abqindustrial net ...

Reviews: