7. Fill up the measuring device with the sample pellets.
The filling needs to be done slowly and constantly to
ensure reproducible results.
8. The display shows the measuring result.
9. To save the results in the storage menu press the
(
button). The storage was successful when the
number in front of the symbol
increased. To reach
the storage menu please press (
) until the
appears.
10. To name the saved results press the
button.
11. Empty the humimeter and ensure that no pellets rests
are accumulated in the measuring chamber.
Measuring range
:
If the measure value is blinking, the valid measuring range
is exceeded (limits see list on page 5). In this case the
accuracy will be decreasing.
Electromatic Equipment Co., Inc.
-13-
11.0 MOST COMMON REASONS FOR MISS READINGS
•
Product temperature out of application range
Material below 0°C resp. above 40°C (32 to 104 °F) may cause faulty
measurements. The storage of cold material in a warm storage area usually
creates condensed water which may lead to major measuring errors.
•
Not adjusted pellets
Let your humimeter BP1 adjust to the surrounding temperature of the material
for approximately half an hour.
A very high temperature difference has a negative effect on the stability of the
measurement result.
•
Wrong calibration curve
Before you measure your sample, double check the correct selection of the
calibration curve.
•
Wrong filling quantity
Fill in exactly 280 gram (± 1.0 g) of wood pellets in the measuring chamber.
•
Wet or mouldy material leads to wrong values!!
•
Frozen measuring material leads to wrong values!!
12.0 DEVICE MAINTENANCE INSTRUCTIONS
To provide a long lives of your device please do not expose it to strong mechanical
loads or heat e.g. dropping it or direct sunlight exposure.
Clean your device using a
dry cloth.
The measuring chamber needs to be cleaned
with a
dry and soft brush
.
Any kind of wet cleaning damages the device. The instrument is not rainproof. Keep
it in dry areas.
We recommend an interval ISO-conforming device check with the drying method.
Electromatic also provides a test with a calibration certificate for a fee.