35
The field of application is limited to industrial frequencies
(15 Hz... 70 Hz)
■
■
■
■
■
Implementation
This function is accessible in AC current measurement only,
after selection of the MIN/MAX mode.
Action
Display
Comments
Press
0.5 P then the value
the yellow
for rms
Enter the function
key
corresponding
out F
Signal frequency
< 15 Hz or > 70 Hz
Press on
1P-2,5P-5P-10P-0,5P
HOLD key
↑
Consultation of values
then press
with each time
rms (computed on
successively
the rms value
of consecutive periods )
the yellow
corresponding
key
alternately
Short pressure Return to values
Exit from the function
on the
MIN, MAX or PEAK
return to MIN/MAX mode
MIN/MAX key
■
■
■
■
■
Characteristics
- Accuracy: 5% +0.5 A
- Capture time: 10 periods of the signal frequency (200 ms
at 50 Hz)
- Range for use:
≥
5 A peak for the first period of the signal
3.8 Power measurement (W)
1. Connect the measurement leads to the instrument’s
terminals, complying with the polarities indicated: red lead
on the “+” terminal and black lead on the “COM” terminal.
2. Set the rotary switch to position “W”.
3. Connect the clamp on the system selected for power
measurement, complying with the following instructions:
- Connect the measurement cables for voltage
measurement, red cable on phase, black cable on neutral.
- Clamp the conductor carrying the current to be measured,
check for proper closing of the jaws and for foreign bodies
in the gap.
The “
➭
“ arrow on the jaws, must be directed in the
direction of power circulation from the source to the load.
In which case:
- the “+” sign corresponds to power consumed by the
load.
- the “-” sign corresponds to power supplied by the load.
■
■
■
■
■
Specific reference conditions
PF = 1; I
≥
2 A; U
≥
10 V