9
10 100 1
10
10
10
100
1
100 1
k
Ω
M
Ω
G
Ω
T
Ω
303.3 M
Ω
3.244 µA
Elapsed Time 00:00:48
DF 20s
984 V
ALARM
DH
I-LIM
The measurement smoothing
filtering is active, with a time
constant of 20 seconds.
Value of the insulation resistance.
True value of the test voltage.
Current flowing between the
terminals.
Elapsed time since the beginning
of the measurement.
Value of the insulation resistance
on the bargraph.
2.2.2. ExAMPLE of dISPLAy dUrING tHE MEASUrEMENt
2.2.3. ExAMPLE of dISPLAy AftEr tHE MEASUrEMENt
995 M
Ω
528 V
531 nA
Elapsed Time 00:01:12
DAR (30s/60s)
PI (1.0m/10m)
Capacitance
1.00
---
2.201 nF
RANGE
E-BRK
ALARM
Value of the insulation resistance.
True value of the test voltage at
the end of the measurement.
The type of measurement is a
non-destructive test.
Current at the end of the
measurement.
Duration of the measurement.
Ancillary results.
The measurement range is fixed.
The voltage generated is >70V
dc
and therefore dangerous.
The
symbol indicates blinking.
If values are undetermined, they are represented by - - - -.
The alarm threshold has been
crossed.
Programmed current limitation.