59
10. Charge the new battery
fully
before using the instrument.
> 90 Vac
< 260 Vac
50 / 60 Hz
11. If the battery was disconnected for more than 5 minutes, you may have to reset the date and time of the instrument (see §5).
Attention:
When the battery is disconnected, even if it is not replaced, it must be fully recharged. This is so that the instrument
will know the battery charge condition (this information is lost when the battery is disconnected).
Attention:
replacing the battery entails the loss of the stored data.
Erase the memory completely (see §6.3.2) to be able to use the MEM/MR functions once again.
9.1.3. rEPLACING tHE fUSE
If the GUARD FUSE message appears on the display unit, the guard terminal fuse must be replaced.
For safety reasons this fuse must always be replaced by an identical model:
Exact type of fuse (marked on the data label): FF - 0,1A - 660V - 6x32mm - 20kA. To order it, see §11.
replacement procedure:
Perform steps 1 to 3 of the battery replacement procedure.