17
CONFIG
ALARM
If the U-FIXED or U-VAR. test voltages have been selected, it is possible to select a measurement configuration by
pressing the CONFIG key (see §4.3):
Manual Stop
Manual Stop + DD
Timed Run
Timed Run + DD
DAR
PI
Then set the type of test, the maximum current, the current range, the filtering of the measurement, and the value
of the alarm threshold:
Test Type
Maximum Output Current
I-range
Disturbance Level
Alarm
DISPLAY
To activate the alarm, press the ALARM key. An audible beep will sound if the result of the measurement is below
the programmed threshold.
3.3.7. bEforE tHE MEASUrEMENt
It is possible to configure the measurement using the CONFIG key
Press the START/STOP button to
start the measurement.
START/STOP
The instrument generates the high voltage. To indicate that the
measurement is in progress, the instrument emits an audible
beep every ten seconds and the START/STOP button lights red.
START/STOP
3.3.8. MEASUrEMENt
At the end of a few seconds, the measurement is
displayed in digital form and in analog form on a
bargraph.
If the measurement is unstable, the instrument automatically filters
the signal, but it is possible to apply a digital filter in addition by
pressing the FILTER key (see §4.6).
FILTER
If the test voltage generated is >5000V, the START/STOP button
blinks.
10 100 1
10
10
10
100
1
100 1
k
Ω
M
Ω
G
Ω
T
Ω
995 M
Ω
528 V
531 nA
Elapsed Time 00:00:23
Set: 500 V
E-BRK