English
45 - 119
4.2.5
Continuity test
Range
Resolution
Uncertainty
Measurement
current
600 Ω
0,1 Ω
Audible signal
< 20 Ω
< 0,35 mA
4.2.6
Diode test
Range
Resolution
Uncertainty
(
±
)
Open-circuit
voltages
Measurement
current
2,8 V
0,001 V
1 % + 2 cts
< 2,8 V
< 0,9 mA
4.2.7
AC/DC Current
600 A AC or DC with clamp having a ratio of 1 mV / 1 A
Range
Resolution
Uncertainty (
±
)
600 A
0,1 A
45 Hz to 1 kHz
2,5 % + 12 cts
(note 1)
Note 1
: not including the uncertainty due to the current clamp
sensor
4.3 ENVIRONMENTAL CONDITIONS
Environmental conditions
In use
In storage
Temperature :
0 °C to +50 °C
-20 °C to +70 °C
Relative humidity (HR) :
<=90% at 40°C
<=50 % at 60 °C
4.4 CHARACTERISTICS OF CONSTRUCTION
Dimension :
H 155 x L 75 x P 55 mm
Weight :
320 g (with the battery)
Measurement
acquisition :
3 times per second
Bargraph :
61 segments, refresh interval 30ms