57
EN
REVOLUTION WIRELESS™
PrOGraM NaMe
VMS-Fr Static 2 cH.
Program Description
The VMS™ FR Static 2 Channels is a physiologically based channel
interaction in which one channel stimulates the agonist and the other
the antagonist of the muscle group that is being exercised. VMS is a
symmetrical biphasic waveform with a 100 μsec interphase interval.
VMS FR 2 Channel delivers current for one Agonist/Antagonist muscle
group. Channel 1 will fire the Agonist while Channel 2 will fire the
Antagonist.
Waveform
VMS-FR
ParaMeterS
SettiNG
Treatment time
30 minutes
Burst Duration
4500 msec
Phase Duration
300µsec
Ramp Time
2 sec
Frequency
35pps
ON/OFF Time
4.5 sec / 13.5 sec
Channels
1+2
Intensity range
1 to 120 mA
Muscle re-education
6 . t r e at M e N t O P t i O N S , F u N c t i O N S a N d d e V i c e S e t t i N G S