Chattanooga
™
Primera Operation Manual
25
Table 202
:
Guidance and manufacturer’s declaration
– electromagnetic immunity
This product is intended for use in the electromagnetic environment speci-
fied below. The customer or the user of this product should assure that it is
used in such an environment, and that precautions regarding that environ-
ment are heeded.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels char-
acteristic of a typical location in
a typical commercial or hospital
environment.