In the event of all 300-Level or a 200-Level error message that cannot be 
resolved, immediately stop all use of the system, and contact the dealer or DJO 
for service. Errors and Warnings in these categories indicate an internal problem 
with the system that must be tested by DJO or a Trained Technician before any 
further operation or use of the system.

-  Use of a system that indicates an Error or Warning in these categories 

may pose a risk of injury to the patient, user, or extensive internal 
damage to the system.

• 

Disconnect the system from the power source before attempting any 
maintenance, installation, removal or replacement procedures to prevent 
electrical shock and possible damage to system.

• 

Vacuum electrodes should not be used on patients with thin, papery skin. 
Vacuum may lead to contact difficulty and bruising.

• 

Vacuum electrodes are not suitable for patients who are taking steroids, due to 
the likelihood of bruising.

• 

Handle, clean and dispose of components and accessories that 
have come in contact with bodily fluids according to National, Local 
and Facility rules, regulations and procedures. 

• 

The solvents of adhesives and flammable solutions used for 
cleaning and disinfecting should be allowed to evaporate before 
the unit is used.

• 

Device is not designed to be used in oxygen rich environment.  
Explosion hazard if the device is used in the presence of flammable 
anesthetic mixture with air, oxygen, or nitrous oxide.

DANGER

WARNING

Summary of Contents for Neo Vacuum Module

Page 1: ...oglobal eu fr_FR index html 2017 DJO LLC All rights reserved NeoVacuum Module Manual 13 8809 Rev E For complete User Operating Instructions including Cautions Warnings Dangers Indications and Contraindications refer to the Intelect Neo ClinicalTherapy System User Manual Item 13 7651 CDVersion item 13 7652 To obtain a copy contact your local representative or DJO Customer Care 0086 ...

Page 2: ... 70006 Vacuum hoses 1 2 Vacuum hoses 3 4 60mm electrodes 60mm Sponges Tools required not included 2 Phillips screwdriver The System is programmed to automatically recognize the new Module therefore no software installation or configuration is required Complete the following steps to install the module 1 Power off device Remove the power cord from the rear of the device 2 Remove the back panel from...

Page 3: ...ach side of the handle Remove the back panel from the Neo Head 4 Insert the Vacuum Module into the Cart as shown Attach the Vacuum Module with the four 2 Phillips screws removed from the back panel in Step 2 5 Thread the Ribbon Cable up to the receptacle on the circuit board plug in and secure with locking clips as shown If needed remove the cable between the power supply and the board prior to se...

Page 4: ... Intelect Neo Clinical Therapy System User Manual item 13 7651 for instructions on installing faceplates 9 Replace the Drawer 10 Reconnect the power cord to the rear of the device MODULE INSTALLATION CONTINUED DONOTpermitforeignmaterials liquidsorcleaningagentstoentertheunit including butnotlimitedto inflammables water andmetallicobjectsfrom enteringtheunit topreventunitdamage malfunction electric...

Page 5: ...ntenance installation removalorreplacementprocedurestoprevent electricalshockandpossibledamagetosystem Vacuumelectrodesshouldnotbeusedonpatientswiththin paperyskin Vacuummayleadtocontactdifficultyandbruising Vacuumelectrodesarenotsuitableforpatientswhoaretakingsteroids dueto thelikelihoodofbruising Handle cleananddisposeofcomponentsandaccessoriesthat havecomeincontactwithbodilyfluidsaccordingtoNat...

Reviews: