Neo Vacuum Module Manual

MODULE INSTALLATION (CONTINUED)

3.  Remove the Back Panel from the Neo Head.

• 

Remove any faceplates from both sides of the cart.

• 

On each side of the cart, remove the screws that 

fasten the back panel to the chassis, as shown.

• 

Remove the handle by removing the two (2) screws 

near the rear of the device from each side of the 

handle.

• 

Remove the back panel from the Neo Head.

4.  Insert the Vacuum Module into the Cart as shown.  

• 

Attach the Vacuum Module with the four #2 Phillips 

screws removed from the back panel in Step 2.

5.  Thread the Ribbon Cable up to the receptacle on the 

circuit board, plug in, and secure with locking clips, as 

shown.

• 

If needed, remove the cable between the power 

supply and the board prior to securing the Vacuum 

Module ribbon cable.  Reattach the power supply 

and board cables after securing the Vacuum Module 

ribbon cable. 

Summary of Contents for Neo Vacuum Module

Page 1: ...oglobal eu fr_FR index html 2017 DJO LLC All rights reserved NeoVacuum Module Manual 13 8809 Rev E For complete User Operating Instructions including Cautions Warnings Dangers Indications and Contraindications refer to the Intelect Neo ClinicalTherapy System User Manual Item 13 7651 CDVersion item 13 7652 To obtain a copy contact your local representative or DJO Customer Care 0086 ...

Page 2: ... 70006 Vacuum hoses 1 2 Vacuum hoses 3 4 60mm electrodes 60mm Sponges Tools required not included 2 Phillips screwdriver The System is programmed to automatically recognize the new Module therefore no software installation or configuration is required Complete the following steps to install the module 1 Power off device Remove the power cord from the rear of the device 2 Remove the back panel from...

Page 3: ...ach side of the handle Remove the back panel from the Neo Head 4 Insert the Vacuum Module into the Cart as shown Attach the Vacuum Module with the four 2 Phillips screws removed from the back panel in Step 2 5 Thread the Ribbon Cable up to the receptacle on the circuit board plug in and secure with locking clips as shown If needed remove the cable between the power supply and the board prior to se...

Page 4: ... Intelect Neo Clinical Therapy System User Manual item 13 7651 for instructions on installing faceplates 9 Replace the Drawer 10 Reconnect the power cord to the rear of the device MODULE INSTALLATION CONTINUED DONOTpermitforeignmaterials liquidsorcleaningagentstoentertheunit including butnotlimitedto inflammables water andmetallicobjectsfrom enteringtheunit topreventunitdamage malfunction electric...

Page 5: ...ntenance installation removalorreplacementprocedurestoprevent electricalshockandpossibledamagetosystem Vacuumelectrodesshouldnotbeusedonpatientswiththin paperyskin Vacuummayleadtocontactdifficultyandbruising Vacuumelectrodesarenotsuitableforpatientswhoaretakingsteroids dueto thelikelihoodofbruising Handle cleananddisposeofcomponentsandaccessoriesthat havecomeincontactwithbodilyfluidsaccordingtoNat...

Reviews: