2. System Set-Up
Caution:
ALWAYS
turn the laser system
OFF
and unplug the power cord from the
wall outlet
before lifting, moving, or relocating the laser system.
2.1. Location
Select a secure and well-ventilated location in which to install and operate the laser system
(see Section 1.3 for description of Laser Treatment Controlled Area).
•
The device should be within 1.8 meters (6 feet) of an available 100-240 V electrical
outlet.
•
Select a flat hard surface that can adequately support the laser system.
•
Ensure adequate airflow around the laser system. The laser system is air-cooled and
designed for use in a well-ventilated location. There must be a minimum 10 cm (4 in)
clearance around the back of the device.
•
An appropriate fire extinguisher should be readily available.
Warning:
Do not operate the laser system in areas with explosion hazards such as
flammable materials or gases.
2.2. Unpacking and Initial Set-up
Step 1.
Carefully unpack the laser system and its accessories from the shipping carton.
Step 2.
Inspect the laser system carefully for any damage, such as cracks, dents or
bent parts.
Step 3.
If items are missing or damaged, call Customer Care. Also, please notify the
carrier if the damage appears to have occurred during shipping and retain
packaging for evidence.
Caution:
If the Warranty Seal Label (
Figure 6
) is not found on the
bottom of the device or is broken,
DO NOT
operate the device. Call
Customer Care.
Step 4.
Fully charge the battery before using this laser system. When
connecting the external power supply to the laser system,
make sure that the cord is correctly aligned. With the flat side of the cord facing
up (
), plug the external power supply into the DC input on the back of
the laser system (
Figure 6