22
Intelect
®
Legend Series Stim
OPERATION
Sweep
is frequency modulation of the High Volt current. When you select
sweep
and press the
enter
button, 4 options are displayed.
They are Continuous (default at 100 pps), 1-10 Hz, 80-150 Hz and 1-150 Hz. The Continuous option allows you to select a continuous
fixed frequency from 1-120 pps.
Frequency
is the number of pulses per second of the waveform. To change the Frequency select
frequency
and press
enter
, then use
the
up
or
down
arrows to change the frequency, from 1-120 pps.
Ramp
controls the amount of time required to bring the stimulation up to the selected
amplitude. When you select
ramp
and press
enter
, 4 options are displayed. They are 0.5
seconds, 1 second, 2 seconds (default) and 5 seconds.
Display
provides two options of viewing output. The options are Voltage (default) and Peak
Current. The ability to assess peak current can help determine tissue response, and an
indication of impedance to current at the electrode skin interface.
Ramp