16
INTELECT
®
RPW 2 QUICK START GUIDE
WARNING
» Increase the separation between the equipment
» Connect the equipment to an outlet on a different circuit from that to which the other device(s)
are connected
» Consult your authorized DJO dealer for help.
• No modification of this equipment is allowed
• The Intelect
®
RPW 2 may be susceptible to Electro-Static Discharge (ESD) at greater than ±8 kV when first
grasping either the Falcon or V actor hand piece. In the event of such a discharge, the Intelect
®
RPW 2 may
experience communication loss. The Intelect
®
RPW 2 will terminate all active outputs, automatically place
the device in a safe state.
• To prevent Electro-Static Discharge (ESD) at greater than ±8 kV:
»
Grasp and hold the hand piece prior to starting treatment. If the applicator must be put down prior to
completion of treatment, stop the current treatment first and then place the applicator in the holder.
»
Maintain humidity in the use environment to at least 35% relative humidity.
»
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, DJO
recommends implementing additional controls to maintain relative humidity to at least 35%.
»
Communicate these ESD-precautionary procedures to healthcare staff, contractors, visitors, and
patients.
• DO NOT block the device's vents.
• Potential electromagnetic or other interference could occur to thsi or other equipment. Try to minimize
this interference by not using other equipment in conjunction with it.
DANGER
• DO NOT connect the device to an electrical supply without first verifying that the power supply is the
correct voltage. Incorrect voltage may cause device damage, malfunction, electrical shock, fire, or personal
injury. Your device was constructed to operate only on the electrical voltage specified on the Voltage
Rating and Serial Number Plate. Contact your DJO dealer if the device is not properly rated.
• Device is not designed to be used in oxygen rich environment, Explosion hazard if the device is used in the
presence of flammable anesthetic mixture with air, oxygen, or nitrous oxide.
• If the device is not safe for operation, it must be repaired by a certified service technician and the operators
must be informed of the dangers posed by the device.
Summary of Contents for Intellect RPW 2
Page 1: ...EN Quick Start Guide ...