Rheometer MDR-01
12
4.2.2) Functions
1. Time Set : is the function that determines the time required to test the
sample (with a range of 0
–
99 minutes) .
For example;
to set the time for 1 minute, just press “1” and then “
E
” to
confirm.
2. Temp Set : is the function that determines the temperature to test the
sample (with a range of 0
–
250 ºC).
For example; to set the temperature for 150ºC, just
press “150” or “150.0”
and then “
E
” to confirm.
3. Preheat time : is setting preheat time function before testing.
Starts from plate is closed. (ranges 0-99 seconds).
Note: The use of this function must be used under the
calibration by an experienced engineer.
4. Stability temp : is the function that configures the operating
functioning temperature while the device is operating. It will provide
alerts on the status of the temperature (ranges from 0
–
9 ºC).
For example; to set the temperat
ure for 1ºC, just press “1” and then
“
E
” to confirm.
5. Time Stability: is setting duration function. When the temperature is
reached the set point on Testing screen, ‘Ready’ will be shown for
notice the user, machine is ready. (ranges 0-999 seconds x 0.1).
For example; to set the time for 1 second
, just press “1”
or 1.0 and
then “
E
” to confirm.
6. Test mode : is testing method function. 2 modes as follows;
1) Time
: Testing time depend on Time Set values.
2) Peak time : Testing time will be stopped when MH values
indicated.