Ceyear 4024 Series Maintenance Manual Download Page 7

Fault cause: The microwave/millimeter-wave frequency conversion module has output fault. 

Instrument to be repaired: Synthetic signal generator 1465H 

Repair method: 

1.  Eliminate the cable connection fault. 

2.  Replace the microwave/millimeter-wave frequency conversion module. 

3.7 SWR deviation at input port 

Fault cause: The RF front end connection has fault or the connection cable has fault 

Repair method: 

1.  Check  if  the  connection  cables  between  the  RF  input  connector,  blocking  module  and 

programmable step attenuator are reliable or loose. 

2.  Please  replace  the  blocking  module  and  the  connecting  semi-rigid  cable  if  the  abnormality  still 

exists. 

3.8 Phase noise deviation 

Fault cause: The IF channel plate or frequency synthesis board has fault. 

Instrument to be repaired: Synthetic signal generator 1465H 

Repair method: 

1.  Replace the IF plate, and verify whether 100 MHz reference ring has any problem or not. 

2.  If the fault is irrelated to the IF plate, it shall be deemed that the frequency synthesis board has local 

oscillator phase noise index deviation, and the frequency synthesis board shall be replaced. 

3.9 Abnormal attenuation control 

Fault cause: The attenuator has communication fault 

Instrument to be repaired: Synthetic signal generator 1465H 

Repair method: 

1.  Check if the communication cable of the attenuator is properly fixed and the metal clip is normally 

installed . 

2.  Replace the rear cover assembly, and check whether the attenuation control is normal. 

3.  Replace the microwave/millimeter-wave frequency conversion module if the fault still exists. 

3.10 

No display for battery under charging and current charging capacity 

Fault cause: The CPU board has fault 

Repair method: 

1.    Maybe the battery is excessive discharged

please charge the battery for half an hour, and see if the 

icon is correct. 

2.    If the display is still incorrect after charged for half an hour, remove the battery and see if the power 

supply icon is the icon of the external power supply. If not, please replace the CPU board. 

 

 

Summary of Contents for 4024 Series

Page 1: ...4024D E F G Spectrum Analyzer 9 kHz 20 GHz 26 5 GHz 32 GHz 44 GHz Maintenance Manual...

Page 2: ......

Page 3: ...dard configuration Multiple measurement function modes Spectrum analysis interference analysis waterfall plot RSSI simulation demodulation AM FM PM channel sweep high precision USB power measurement e...

Page 4: ...ency band are mixed with the first LO signal after the preamplifier and subsection switching filter to produce first IF signal 5 GHz 16 5 GHz The first IF signal and RF signals of 9kHz 20 GHz are swit...

Page 5: ...as the AD converter input sampling signal After AD sampling it will output the LVDS level to the data processing display and control module for IF filtering video detection digital down conversion and...

Page 6: ...ctly connected 2 If the cable is normally connected please replace the frequency synthesis board and power it on and check it for zero frequency 3 Please replace the microwave millimeter wave frequenc...

Page 7: ...r not 2 If the fault is irrelated to the IF plate it shall be deemed that the frequency synthesis board has local oscillator phase noise index deviation and the frequency synthesis board shall be repl...

Page 8: ...mposition and remove and install it according to the instruction 4024 Exploded View Interface protection cover LCD touch screen Keyboard Fan assembly CPU board Battery compartment Frequency synthesis...

Page 9: ...e casing with a Phillips screwdriver Fig 3 Removal of 6 Fixing Screws 3 Separate 4024 front casing from its rear casing Attention In this case two parts of the casing are connected by multiple cables...

Page 10: ...rews of the frequency synthesis board 12 in the rear cover assembly 19 b Loosen 3 semi rigid cables between the frequency synthesis board 12 and the microwave frequency conversion module 13 c Remove t...

Page 11: ...Remove the connection flat cable between the attenuator 15 and the microwave frequency conversion module 13 c Remove the connection semi rigid cable between the attenuator 15 and the microwave frequen...

Page 12: ...PU board 9 a Remove the interconnecting 50 core cables between the CPU board 9 and the keyboard 5 b Remove the connection cables between the CPU board 9 and 2 fan components 8 c Remove 4 fixing screws...

Page 13: ...strument name Main technical indicators Recommended model Manufacturer Synthesized signal generator Frequency range 250 kHz 50 GHz Power output 100 dBm 15 dBm Frequency accuracy 0 02 The power level c...

Page 14: ...op up As shown in the following figure the GPIB address of the signal source to be connected the GPIB Address of the power meter and the IP Address of the spectrum analyzer can be set The address of t...

Page 15: ...e cables The interface is as shown in the following figure When the test is completed it will prompt to generate the file PowerMeter powdoc Fig 12 Generate Power Meter File d This step is mainly inten...

Page 16: ...t within the same network segment for example if the IP of the spectrometer is 172 141 11 202 the IP of the computer shall be set to 172 141 11 xx Fig 14 Measure Spectrum Analyzer Connect the computer...

Page 17: ...nd low frequency output function 1465H with 90 dB programmable step attenuator option Ceyear Power meter Frequency range 50MHz 67GHz Power range 60 dBm 20 dBm 2438CB Ceyear Power sensor Frequency rang...

Page 18: ...as shown in the following figure Fig 15 Schematic Diagram of Connection of the Automatic Test 2 Open 4024AutoTest exe and the interface is as follows Fig 16 Interface of the Automatic Test of Indexes...

Page 19: ...Test Attention In case of measurement and display of the average noise level and remaining response a 50 load shall be connected at the RF port The indicators including test frequency range third ord...

Reviews: