cable with the PC through the GPIB card, and connect 4024 to be tested with the PC through the
direct-connected network cable. It is as shown in the figure:
Fig. 9 Schematic Diagram of Connection
b) Open the temperature compensation software TempComp.exe, and the software interface is as
follows:
Fig. 10 TempComp Interface
Click the Set Address button, and the communication address setting dialog will pop up. As shown in the
following figure, the GPIB address of the signal source to be connected, the GPIB Address of the power
meter and the IP Address of the spectrum analyzer can be set. The address of the signal source and
power meter can be inquired through
【
System
】
menu. If the GPIB number of the signal source is 19,
no modification is required; similarly, if the GPIB number of the power meter is 13, no modification is
required. The IP address of the spectrometer can be inquired through
【
System
】
->[Network Config].
GPIB cable
Signal source
Adapter
2.4 mm cable
PC
Adapter
Power
sensor
Power meter
GPIB card