Ceyear 4024 Series Maintenance Manual Download Page 17

after  the  measurement  is  completed,  as  shown  in  the  following  figure.  When  this  file  and 
“IFGainAtten.dc” are in the same folder, the temperature compensation is completed. 

5.2 Automatic Test of Indicators 

Software environment 

Name of the software 

Version 

Description 

Windows system 

Win7/XP 

The compensation software operating environment needs 
GPIB driver and Visa library 

4024AutoTest.exe 

1.0.5 

Software for automatic test of indicator 

NI-GPIB driver.rar 

GPIB-USB-HS 

NI GPIB driver installation package 

visa462full.exe 

4.6.2 

Visa library installation package 

Environment 

Instrument name 

Main technical indicators 

Recommended model 

Manufacturer 

Synthesized signal 

generator 

Frequency range: 250 kHz ~ 50 

GHz 

Power output: -100 dBm ~ +15 dBm 

Frequency accuracy: ±0.02

 

The power level can be calibrated 

and stored 

It has internal and external AM 

modes and low frequency output 

function 

1465H (with 90 dB 

programmable step 

attenuator option) 

Ceyear 

Power meter 

Frequency range: 50MHz ~ 67GHz 

Power range: -60 dBm ~ +20 dBm 

2438CB 

Ceyear

 

Power sensor 

Frequency range: 50MHz ~ 67GHz 

Power range: -60 dBm ~ +20 dBm 

71710L 

Ceyear

 

Low-pass filter 

Cutoff frequency: 1.0 GHz, insertion 

loss<0.9 dB, off-band 

suppression>65 dB 

SLP-1200 

Mini-Circuits 

Low-pass filter 

Cutoff frequency: 6.7GHz, insertion 

loss<1 dB, off-band suppression>40 

dB 

VLF-6700 

Mini-Circuits 

Low-pass filter 

Cutoff frequency: 18GHz, insertion 

loss<2 dB, off-band suppression>40 

dB 

FLP-1800 

Marki 

50Ω adapter 

Impedance: 50Ω, interface type: 2.4 

mm female 

70508 

Ceyear

 

Adapter 

2.4 mm (f) - 2.4 mm (f) 

71120ME 

Ceyear

 

Adapter 

2.4mm(f)-3.5mm(f) 

71125ME 

Ceyear

 

Adapter 

2.4 mm (m) - 3.5 mm (f) 

71123ME 

Ceyear

 

Adapter 

2.4 mm (f) - N (m) 

81107A 

Ceyear

 

Adapter 

2.4 mm (m) - N (m) 

81107C 

Ceyear

 

Cable 

BNC (m) - BNC (m), 2 m 

BNC-JJ-2000mm 

Ceyear

 

Low loss cable 

2.4 mm cable (m-m), 1.5 m 

2.4 mm-JJ-1500 mm

 

Ceyear

 

USB-GPIB control 

card 

USB-GPIB control card 

GPIB-USB-HS 

NI 

GPIB cable 

1.5 m 

GPIB cable, 1.5 m 

Zetai 

Summary of Contents for 4024 Series

Page 1: ...4024D E F G Spectrum Analyzer 9 kHz 20 GHz 26 5 GHz 32 GHz 44 GHz Maintenance Manual...

Page 2: ......

Page 3: ...dard configuration Multiple measurement function modes Spectrum analysis interference analysis waterfall plot RSSI simulation demodulation AM FM PM channel sweep high precision USB power measurement e...

Page 4: ...ency band are mixed with the first LO signal after the preamplifier and subsection switching filter to produce first IF signal 5 GHz 16 5 GHz The first IF signal and RF signals of 9kHz 20 GHz are swit...

Page 5: ...as the AD converter input sampling signal After AD sampling it will output the LVDS level to the data processing display and control module for IF filtering video detection digital down conversion and...

Page 6: ...ctly connected 2 If the cable is normally connected please replace the frequency synthesis board and power it on and check it for zero frequency 3 Please replace the microwave millimeter wave frequenc...

Page 7: ...r not 2 If the fault is irrelated to the IF plate it shall be deemed that the frequency synthesis board has local oscillator phase noise index deviation and the frequency synthesis board shall be repl...

Page 8: ...mposition and remove and install it according to the instruction 4024 Exploded View Interface protection cover LCD touch screen Keyboard Fan assembly CPU board Battery compartment Frequency synthesis...

Page 9: ...e casing with a Phillips screwdriver Fig 3 Removal of 6 Fixing Screws 3 Separate 4024 front casing from its rear casing Attention In this case two parts of the casing are connected by multiple cables...

Page 10: ...rews of the frequency synthesis board 12 in the rear cover assembly 19 b Loosen 3 semi rigid cables between the frequency synthesis board 12 and the microwave frequency conversion module 13 c Remove t...

Page 11: ...Remove the connection flat cable between the attenuator 15 and the microwave frequency conversion module 13 c Remove the connection semi rigid cable between the attenuator 15 and the microwave frequen...

Page 12: ...PU board 9 a Remove the interconnecting 50 core cables between the CPU board 9 and the keyboard 5 b Remove the connection cables between the CPU board 9 and 2 fan components 8 c Remove 4 fixing screws...

Page 13: ...strument name Main technical indicators Recommended model Manufacturer Synthesized signal generator Frequency range 250 kHz 50 GHz Power output 100 dBm 15 dBm Frequency accuracy 0 02 The power level c...

Page 14: ...op up As shown in the following figure the GPIB address of the signal source to be connected the GPIB Address of the power meter and the IP Address of the spectrum analyzer can be set The address of t...

Page 15: ...e cables The interface is as shown in the following figure When the test is completed it will prompt to generate the file PowerMeter powdoc Fig 12 Generate Power Meter File d This step is mainly inten...

Page 16: ...t within the same network segment for example if the IP of the spectrometer is 172 141 11 202 the IP of the computer shall be set to 172 141 11 xx Fig 14 Measure Spectrum Analyzer Connect the computer...

Page 17: ...nd low frequency output function 1465H with 90 dB programmable step attenuator option Ceyear Power meter Frequency range 50MHz 67GHz Power range 60 dBm 20 dBm 2438CB Ceyear Power sensor Frequency rang...

Page 18: ...as shown in the following figure Fig 15 Schematic Diagram of Connection of the Automatic Test 2 Open 4024AutoTest exe and the interface is as follows Fig 16 Interface of the Automatic Test of Indexes...

Page 19: ...Test Attention In case of measurement and display of the average noise level and remaining response a 50 load shall be connected at the RF port The indicators including test frequency range third ord...

Reviews: